A-535
8
E N G L I S H
8
O P E R AT I O N
3. Inspect the oxygen fill connector for the following:
a) Debris, oil or grease
b) Noticeable signs of damage
c) Signs of corrosion inside the valve
d) Signs of excessive heat or fire damage
Do not use oxygen cylinder if found, call your homecare provider.
OPERATION
CONNECTING THE IFILL OXYGEN CYLINDER TO THE DEVILBISS IFILL
PERSONAL OXYGEN STATION
DANGER
Fire Hazard. Do not lubricate. Do not allow grease or oil from your hands or other
source to come into contact with the regulator or cylinder valve connection. These
solutions may be flammable and cause injury.
WARNING
Do not modify ANY connections on the personal oxygen station. NEVER
use tools of any kind to connect/disconnect the cylinder and the oxygen station.
Severe injury and/or damage may occur.
Do not drop oxygen cylinders. Use two (2) hands when handling/transporting
oxygen cylinders. Otherwise, injury or damage may occur.
NOTE
–
Refer to the G
ENERAL
D
ANGERS
& W
ARNINGS
and the H
ANDLING
W
ARNINGS
in this manual.
1. Plug the power cord into the wall outlet.
2. Set the rotary selector on the cylinder to “OFF.”
3. Remove the oxygen fill connector cover from the fill connector.
4. Position the cylinder over the cradle while aligning the nipple connector on the oxygen
cylinder with the fill connector. Press until cylinder “Clicks” into place (Figure 2).
NOTE
–
When using the smaller M4 or M6 size cylinders, it is necessary to place the oxy-
gen fill connector cover/cylinder adapter in the cradle to help support the cylinder. When
using the larger ML6, C, D, or E cylinders, the oxygen fill connector cover/cylinder adapter is
not necessary and can be allowed to hang over the side.
FILLING THE IFILL OXYGEN CYLINDER
NOTE
–
Refer to the G
ENERAL
D
ANGERS
& W
ARNINGS
and the H
ANDLING
W
ARNINGS
in this manual.
1. Make sure the personal oxygen station is plugged in.
2. Make sure the cylinder is securely connected to the oxygen station. Refer to
Connecting
The iFill Oxygen Cylinder To The DeVilbiss iFill Personal Oxygen Station.
3. Push the start/stop switch on the control panel (Figure 3).
4. The following sequence of events should occur:
a. The FILLING (green) light will illuminate while the cylinder is filling (refer to
iFill
Oxygen Cylinder Fill Times
).
b. The FULL (green) light will illuminate when the cylinder is full. Proceed to
Removing
the iFill Oxygen Cylinder
.
NOTE
–
If the Service (RED) light and audible alert are on, proceed to T
ROUBLESHOOTING
in this
instruction manual.
INDICATOR LIGHT EXPLANATION
INDICATOR LIGHT COLOR
OXYGEN STATION STATUS
EFFECT
None
Oxygen station is unplugged.
None
Standby (green)
Oxygen station is ready to begin filling cylinders.
Cylinder filling not started
Filling (green)
Oxygen station is on and cylinder is filling.
Cylinder is filling
Full (green)
Cylinder is full.
Remove cylinder.
Service (red) and Audible Alert
Oxygen station is on but cylinder is not filling due to inter-
nal failure.
Contact your Sunrise Medical
Provider
Figure 2
Figure 3