A-535
10
E N G L I S H
10
T R O U B L E S H O O T I N G / C L E A N I N G / M A I N T E N A N C E
IFILL OXYGEN CYLINDER/REGULATOR
TYPICAL QUESTIONS AND ANSWERS
Q.
Why does my iFill make a “pop” sound when I remove the cylinder?
A.
The “pop” sound is pressure being released from the system to make the next cylinder easier to install.
Q.
Why is the coupler only to be cleaned with a clean, dry, lint-free cloth and only when necessary?
A.
Cleaning solutions may be flammable and/or may contaminate the cylinder.
Q.
Why can’t I use my old cylinders with the iFill?
A.
The iFill has a proprietary connection as required by the FDA and, therefore, other cylinders are not compatible.
CLEANING/MAINTENANCE
DEVILBISS IFILL PERSONAL OXYGEN STATION FILTER
CAUTION
–
Do not operate the personal oxygen station without the filters installed.
NOTE
–
The DeVilbiss iFill Personal Oxygen Station has two (2) filters, one on each side of the cabinet (FIGURE A).
NOTE
–
Perform this procedure at least ONCE A WEEK.
1. Unplug the personal oxygen station from the wall outlet.
2. Remove the filters.
3. Clean the filters with a vacuum cleaner or wash in warm soapy water and rinse thoroughly.
4. Dry the filters thoroughly before reinstallation.
EXTERIOR CABINET
WARNING
Electric Shock Hazard: Unplug the DeVilbiss iFill Personal Oxygen Station when cleaning. Do not remove the oxygen
station cabinet. The cabinet should only be removed by a qualified Sunrise Medical/DeVilbiss technician. Do not apply
liquid directly to the cabinet nor utilize any petroleum-based solvents or cleaning agents.
1. Clean the exterior cabinet by using a damp cloth or sponge with a mild household cleaner and wipe it dry.
2. Only if necessary, wipe the coupler with a clean, dry, lint-free cloth.
PROBLEM
SOLUTION
Hissing sound from iFill oxygen cylin-
der
1. Turn rotary selector OFF.
2. If hissing sound continues, call your Sunrise Medical/DeVilbiss provider.
iFill oxygen cylinder is on but oxygen
is not being delivered.
1. Refill cylinder.
2. Check that cannula is straight and not pinched. If damaged, replace.
3. Call your Sunrise Medical/DeVilbiss provider for service.