EN - 11
SE-DV57-1
SUPPLEMENTAL OXYGEN
WARNING-OXYGEN USE
•
Oxygen supports combustion. To avoid possible physical injury, do not smoke while using this device with
supplemental oxygen. Do not use this device near hot objects, volatile substances or sources of open flames.
•
Always turn on the SleepCube before turning on the oxygen source. Turn off the oxygen source before turning off the
SleepCube.
•
This device is not suitable for use in the presence of a flammable anesthetic mixture.
•
At a fixed flow rate of supplemental oxygen, the inhaled oxygen concentration will vary depending on the pressure
setting, patient breathing pattern, mask selection and leak rate. This warning applies to most types of CPAP devices.
•
Never allow the oxygen source to run continuously while connected to the SleepCube if the device is not in use. If
the SleepCube is not being used, turn off the oxygen flow.
CAUTION
•
Oxygen is a prescribed gas and should only be administered under the supervision of a physician.
•
The flow setting of the oxygen source must be specified by a physician.
Supplemental oxygen is sometimes prescribed in addition to a CPAP for treatment of Obstructive Sleep Apnea. If your
physician prescribed oxygen for your sleep therapy, you can add the oxygen in one of two ways:
1. Obtain the optional oxygen adapter (part #7353D-601) from your homecare provider and connect it to the air supply
port on the SleepCube. Connect one end of the air supply tubing to the open end of the adapter and the other end of
the air supply tubing to the mask.
2. If your mask is equipped with an oxygen port, your physician may request that you apply oxygen directly to your mask.
ACCESSORIES/REPLACEMENT ITEMS
Replacement Items
Air-inlet filter (4/pk)
DV51D-602
Heater connection cover
DV51D-605
Fine particle filter (4/pk)
DV51D-603
Oxygen adapter
7353D-601
Air supply tubing (6’)
CPAPTUB6-90 Carrying case
DV51D-610
Air supply plug
DV51D-604
Heated Humidifier
DV5HH
Humidifier Chamber
DV5C
Chamber Sealing Gasket
DV5C-614
Power Items
AC power cord (USA)
DV51D-606
AC power cord (Australia)
DV51D-609
AC power cord (UK)
DV51D-608
DC power cord
DV51D-619
AC power cord (EU)
DV51D-607
DC battery clamp-on adapter (requires DC
Power Cord DV51D-619)
DV51D-696
MAINTENANCE
Expected Service Life:
•
CPAP and humidifier cradle = 5 years
•
Humidifier water chamber = 1 Year
•
Tubing and gross particle intake filter = 180 days
DANGER
Electric shock hazard – Do not attempt to open or remove the cabinet. There are no user-serviceable components
inside. If service is required, contact your equipment provider for instructions on obtaining service. Opening or
attempting to service your device will void the warranty.
WARNING
Disconnect supply before servicing.
Pressure Accuracy–The SleepCube is designed to provide accurate pressure performance from the factory without
additional calibration. No routine service is required provided the device is used in accordance with the manufacturer’s
directions.