EN - 10
SE-DV57-1
Mask Off – This message appears due to a poor mask fit or a removed mask. Check for leak around the mask seal and
make adjustments as necessary according to the mask manufacturer’s instructions. This message will be displayed until
the high airflow problem is corrected. If the high airflow condition persists for about 20 seconds, the device will
automatically turn off if Auto-OFF is enabled.
Device Fault – If a device fault message is displayed, refer to Troubleshooting for instructions.
SLEEPCUBE TRAVEL INFORMATION
International Power Changes
Your DeVilbiss SleepCube is equipped with a universal power supply. This means that it is automatically capable of
accepting line voltages of 100 to 240V~ 50/60Hz. You do not need to make any adjustments to the device if you are
traveling to another country where the line voltage is different than your domestic line voltage. Simply contact your
equipment provider for the correct style power cord for the region in which you will be traveling (refer to Accessories/
Replacement Items for the correct part number).
Altitude
Your SleepCube automatically compensates for changes in altitude between sea level and 8500 ft. (2600 m). You do not
need to make adjustments to the device for changes in altitude.
SLEEPCUBE DC OPERATION
Battery Power
our SleepCube can be operated from a 12V DC power source in a motor home, boat or other recreational vehicle using
the 12V DC cable. Alternatively you can power your SleepCube using a 12V DC stand-alone (deep-cycle marine) battery
with a DC battery clamp on adapter and 12V DC cable. All cables are available from your equipment provider (refer to
Accessories/Replacement Items for the correct part numbers). The stand-alone battery connection uses clamps that attach
to the terminal connections of the battery.
Approximate Battery Run Time (with fully charged battery) @ 12 cm H
2
O setting
SleepCube Connected To Vehicle
SleepCube Connected To Battery
DC Power
Source/Battery
DC Car Power Source (Cigarette Lighter)
Deep Cycle Marine Battery
Battery Specs
12 V
12V
12V
80 Amp-Hours 120 Amp Hour 240 Amp Hour
Additional
Equipment Needed
120V~
Modified Sine
Wave Inverter
with 200 Watt
capacity
230V~ Modified
Sine Wave
Inverter with
400 Watt
Capacity
120 VAC
Modified Sine
Wave Inverter
with 200 Watt
capacity
230 VAC
Modified Sine
Wave Inverter
with 400 Watt
Capacity
DeVilbiss Cables
Required (refer to
Accessories)
DV51D-619 DV51D-696
DV15D-696
DV51D-619
DV51D-696
DV51D-696
DV51D-696
Heater Humidifier
Powered?
NO
YES
YES
NO
YES
YES
Approximate Run
Time
5 nights
1 night
1 night
7 nights
1 night
1 night
NOTE–Run times are at a breath rate of 20 breaths per minute, standard mask leak, and with heater setting of 8 where
applicable.
NOTE–The heater for the SleepCube’s humidification system will not operate directly from a 12V DC power source. If you
need humidification, non-heated humidification is still available. Alternately, instead of connecting12V DC power directly to
the SleepCube, you can pass the DC power through an inverter that will deliver AC power to the system. This will allow the
heater to function. The power rating of the inverter must be at least 200 watts @ 110V~ / 400 watts @ 220V~.
NOTE–The SleepCube will automatically switch power sources between AC and 12V DC if both sources are available. If
your unit is connected to both AC and 12V DC power sources, it will operate on AC power as long as AC power is present
and automatically switch to 12V DC power if AC power is lost. When AC power is restored, it will return to the AC source.
You do not need to reset or adjust any settings when the power source switches, but remember that your heater will not
operate on 12V DC.