38
15.
COMPATIBILITY & ACCESSORIES
Refer to Drive DeVilbiss Healthcare Ltd. bed instructions for use to ascertain
suitable mattress compatibility.
The Artemis II control unit is suitable for use with:
• Artemis II Dynamic Mattress (sold as system only) – DYN/DIG/ARTEMIS/2
• Artemis II Dynamic Cushions – ARTEMIS/2/CUSH/18 and ARTEMIS/2/CUSH/20
Ensure the mattress/cushion is used with a compatible side rail,
bed frame, or chair, combination – incorrect combinations can lead
to entrapment and/or falls hazards.
Warning
15.1
Artemis II Cushions
The Artemis II 18” and 20” cushions are intended for chairs which have a platform
of approximately 18” x 18”, or 20” x 20” respectively.
15.2
Artemis II Mattress
The Artemis II mattress is intended for bed frames which have a platform of
approximately 2000mm x 900mm, however the choice of side rail used can
determine suitable compatibility or not.
Drive DeVilbiss Healthcare bedframes reference suitable mattress compatibility
in their instructions for use along with side rail compatibility. These instructions
are to be consulted prior to using a mattress with a specific bed and side rail
combination. Alternatively contact Drive DeVilbiss Healthcare or the local
distributor for compatibility advise. If in doubt always seek advice.
Immunity test
IEC 60601 test level
Compliance level
Conducted RF
IEC 61000- 4-6
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms:
in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms:
in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
Radiated RF
IEC 61000- 4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz