13.1 ARTEMIS II MATTRESS SPECIFICATION
Classification:
Electrical shock protection: Class II, Type BF
Applied Part: Mattress
Ingress protection:
IP21***
Not AP or APG equipment*
Supply Rating:
230V, 50Hz, 25W
Fuse Rating:
Mains Plug – 10A
Mains plug:
Type G/BS1363
Battery Type:
NiMH (Nickel Metal Hydride)
Mattress Dimensions:
(L) 2000mm x (W) 880mm x (D) 200mm
Mattress Weight:
11kg
Maximum Patient Weight:
267kg (42 stone)
No. of Cells:
23 cells which include:
3 static head cells
20 alternating cells (inc. 8 narrow heel cells)
Alternating Therapy:
AB pattern
Cycle Time:
10, 15 or 20 minutes
Pressure Range:
Alternating Mode: 13 – 70 ±2mmHg
Constant Low Pressure Mode: 10 – 40 ±2mmHg
Pulsation Mode: +30% & -30% of CLP setting
Max Inflate: 70 ±2mmHg
Control Unit Dimensions:
(L) 180mm x (W) 390mm x (D) 165mm
Control Unit Weight:
4.1 kg
Cover Material:
Dartex®
Cell Material:
TPU
Base Material:
Nylon/PU
Transport and Storage Conditions:
Ambient Temp: -25°C to +70°C
Humidity: < 93%, non-condensing
Operational Conditions:
Ambient Temp: +5°C to +40°C
Humidity 15% - 93%, non-condensing
Atmospheric Pressure:
700hPa to 1060hPa
Operating Altitude:
≤ 2000m
Pollution:
Degree 2
UV:
Intended for indoor use only
Noise level:
<40dB(A)
Safety Standards:
IEC/EN 60601-1
IEC/EN 60601-1-11
IEC/EN 60601-1-2
Cover complies with BS7175:1989 - Medium Hazard
Expected Service Life: