31
Incomplete
inflation / Low
pressure
•
Audible signal
sounds.
•
Screen shows:
1.
Ensure the mattress air connector is correctly
connected to the control unit.
2. Ensure the CPR dial is closed and there is no air
leakage.
3. Turn the unit off and then on again to clear the
indicator.
If a ‘low pressure’ indicator continues to illuminate:
4. Open the mattress and ensure there is no air
leakage within the mattress – cells, tubing and
connectors.
5. Turn the unit off and then on again to clear the
indicator.
If a low pressure indicator is still evident turn off at the
mains and contact your approved service provider.
High Pressure
•
Audible signal
sounds.
•
Screen shows:
1.
Ensure the mattress umbilical is not trapped or
being squeezed.
2. Open the mattress and ensure none of the air
pipes are kinked.
3. Turn the unit off and then on again to clear the
indicator.
If a high pressure indicator is still evident turn off at the
mains and contact your approved service provider.
Mattress
Disconnection
•
Audible signal
sounds.
•
Screen shows:
1.
Ensure the mattress air connector is correctly
connected to the control unit.
2. Turn the unit off and then on again to clear the
indicator.
If the indicator is still evident turn off at the mains and
contact your approved service provider.
Patient is
bottoming out
1.
Ensure the patient is suited to the maximum
rating of the mattress.
2. Ensure the patient is centrally positioned on the
mattress.
3. Increase the pressure setting – Refer to section
9.5.4.2
Egress Alert
•
Audible signal
sounds.
1.
Ensure the patient is centrally positioned on the
mattress.
2. Turn the unit off and then on again to clear the
indicator.
*** LOW PRESSURE ***
Please check system
including CPR. To restart Artemis II press
Power off/on.
*** HIGH PRESSURE ***
*** SYSTEM FAILURE ***
Please check system
including CPR. To restart Artemis II press
Power off/on.