Artemis II Dynamic Mattress and Cushions

Instructions for use

Содержание drive Artemis II

Страница 1: ...Artemis II Dynamic Mattress and Cushions Instructions for use...

Страница 2: ...button FIG 4 and pulling away FIG 5 Switch off the control unit CPR mattress only Rapid deflation of the mattress may be required for emergency treatment or system deflation The CPR dial is located a...

Страница 3: ...es if power is lost 7 MAX INFLATE MODE COMFORT CONTROL Opens comfort control menu 1 Press MAX INFLATE MODE COMFORT CONTROL to open Comfort Control Menu 2 Press down MAX INFLATE MODE COMFORT CONTROL MU...

Страница 4: ...ng Mode 20 9 5 2 2 Patient Weight Pressure 21 9 5 3 Safety Functions 21 9 5 3 1 Function Lock 22 9 5 3 2 Mute 22 9 5 4 System Functions 22 9 5 4 1 Mode Selection 22 9 5 4 2 Max Inflate 23 9 5 4 3 Comf...

Страница 5: ...DI and serial numbers can be found on the back of the control unit and inside the mattress cushion For Service Support outside the UK Northern Ireland please contact the local distribution company fro...

Страница 6: ...n overall programme of care when active load distribution through mechanical means is required 3 3 Intended Use The intended use of the mattress is to support the weight of the patient as identified w...

Страница 7: ...e to a pre determined value but the comfort level can be adjusted by manually adjusting the cell pressure up or down Should a fault occur such as a power failure or loss of pressure an audio visual al...

Страница 8: ...lity of the patient to the mattress size Patient with burns Unauthorised people with access to the controls Patients who have reduced capacity and are agitated and or restless Small adults children Su...

Страница 9: ...aters away from the support surface Follow all manufacturers instructions and warnings It is advised that a full fire risk assessment is carried out prior to using this equipment In case of fire exit...

Страница 10: ...n fire Control unit functions must be locked out when a patient is left unattended If children adults who lack capacity or even pets pose a potential risk of intentional or unintentional tampering wit...

Страница 11: ...torage ensure it has been cleaned and disinfected in line with the local infection control policy and or as defined in section 10 of these instructions for use Do not remove the mattress cushion from...

Страница 12: ...ot dry clean Tumble dry on low heat Zip location Keep out of direct sunlight Max patient weight Medical device MD SN Serial Number No Smoking Temperature limit Class II Electrical Device The user is p...

Страница 13: ...parts of the device that come into contact with the patient in order to carry out its intended function refer to section 16 2 Type BF Applied parts which are electrically isolated from earth and other...

Страница 14: ...ication If the system has come from a storage transport temperature environment near to the minimum or maximum values stated allow the cushion mattress and control unit to adjust to room temperature f...

Страница 15: ...raps For profiling beds it is essential that the straps are secured around the movable sections of the bed frame Damage will be incurred when profiled if secured to fixed parts of the frame To avoid a...

Страница 16: ...ttress specification could lead to an injury The mains plug is the disconnect device for the means of isolating the control unit from the mains supply the plug must be accessible at all times Ensure t...

Страница 17: ...oisture could damage the electrical system and pose an electrical fire risk Ensure the cushion is used with a compatibly sized chair incorrect combinations can lead to entrapments falls hazards Incomp...

Страница 18: ...menu 8 Mode select 9 3 Control Interface No Symbol Description 9 Enters maximum inflation 10 Returns to Home screen No Symbol Description 1 Turns system on off 2 Mutes the audible signal for a 20 minu...

Страница 19: ...Cycle Time UP DOWN KEY 4 Extended Functions Page 2 2 Comfort Control Menu Press Enter Button to Enable Selection Home Enter Key Mode Key Comfort Control Key Home Enter Key Mode Key Home Screen Comfort...

Страница 20: ...r 45 minutes see section 11 To turn the system off hold the button down for 2 seconds until the screen extinguishes Once a patient is positioned on the mattress the system will determine the patient s...

Страница 21: ...t indicator is not to be used for medical purposes it is to act as information only If the inclusion of the weight indicator could introduce a hazard it can be hidden see 9 5 4 3 Warning Once the syst...

Страница 22: ...ng When pressed a menu appears offering 3 different settings 9 5 4 System Functions 9 5 4 1 Mode Selection 1 Alternating 2 Constant Low Pressure 3 Pulsation Use the up down cursor key followed by the...

Страница 23: ...fety Before changing the pressure or cycle time clinical judgement is required from frequent monitoring and repositioning of the patient 9 5 4 3 Comfort Control 9 5 4 4 Comfort Level When pressed a co...

Страница 24: ...3 Auto Lock a On b Off When pressed a menu appears offering 2 different settings 4 Weight Unit a kg b lbs By selecting this function the cycle time of the alternation sequence can be manually altered...

Страница 25: ...activate the egress alert If active and the occupant gets out of bed an audible signal sounds and a warning screen illuminates for approximately 1 minute followed by a no patient indication To cancel...

Страница 26: ...t your service engineer Service Due Service due contact maintenance provider as soon as possible 9 6 2 Cushion Mode The seat cushion can be operated with the Artemis II pump the system will be automat...

Страница 27: ...lling away from the control unit 2 9 8 Support Surface Disconnection and Power Cuts 1 2 Switch off the control unit if still operational Disconnect from the power supply Note on the mattress there is...

Страница 28: ...cleaning must be carried out in accordance with your local infection control policy or regulatory body Warning Always disconnect the support surface and bed frame from the main power supply prior to...

Страница 29: ...de Any evidence of strike through and or cover damage will require a new cover to be fitted to the mattress Warning The cover must not be used if strike through is evident Risk of cross infection Freq...

Страница 30: ...pproved service engineer 1 Turn off the control unit to silence the alert and turn off the mains supply Note the mute button does not silence the power failure indication 2 Check the mains cable is fu...

Страница 31: ...n on again to clear the indicator If a high pressure indicator is still evident turn off at the mains and contact your approved service provider Mattress Disconnection Audible signal sounds Screen sho...

Страница 32: ...imum Drive DeVilbiss Healthcare Ltd also recommends that the carer performs frequent visual and operational inspections If there are any signs of damage or the system is not performing as it should wi...

Страница 33: ...n alternatively follow local recycling and disposal policies The control unit used with your system is not to be disposed of in general municipal waste as it is to be considered as Waste Electrical an...

Страница 34: ...0 15 or 20 minutes Pressure Range Alternating Mode 13 70 2mmHg Constant Low Pressure Mode 10 40 2mmHg Pulsation Mode 30 30 of CLP setting Max Inflate 70 2mmHg Control Unit Dimensions L 180mm x W 390mm...

Страница 35: ...0 15 or 20 minutes Pressure Range Alternating Mode 40 90 2mmHg Constant Low Pressure Mode 10 40 2mmHg Pulsation Mode 30 30 of CLP setting Max Inflate 90 2mmHg Control Unit Dimensions L 180mm x W 390mm...

Страница 36: ...to operate abnormally turn off at the mains supply and contact Drive DeVilbiss Healthcare Ltd or your local distributor see section 2 Portable RF communications equipment e g mobile cordless phones s...

Страница 37: ...vel Compliance level Electrostatic discharge ESD IEC 61000 4 2 8 kV contact 15 kV air 8 kV contact 15 kV air Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 2kV for power sup...

Страница 38: ...bed frames which have a platform of approximately 2000mm x 900mm however the choice of side rail used can determine suitable compatibility or not Drive DeVilbiss Healthcare bedframes reference suitabl...

Страница 39: ...iled by or attributes to acts of God f damaged caused by failure to follow instructions or g otherwise used in a manner inconsistent with any instructions provided by Drive DeVilbiss Healthcare Ltd Th...

Страница 40: ...vedevilbiss co uk www drivedevilbiss co uk Drive DeVilbiss Healthcare Ltd Sidhil Industrial Estate Holmfield Halifax HX2 9TN West Yorkshire United Kingdom A member of the Drive DeVilbiss Group of Comp...