neoprobe
TM
GDS Operation Manual
AW-000639
Page
6 of 38
Quick Start BLUETOOTH
®
Wireless Technology Enabled Probe
Step
Prepare probe equipped with BLUETOOTH
®
wireless technology
1
Prepare for operation as stated above
2
Install battery in probe
3
Press button on probe; probe blue light is on steady
4
Press and hold Target Count Button [11] on console; displays “disc”; release Target Count
button
5
Probe connects; blue light on probe flashes and starts normal operation
See manual provided with probe for complete instructions
When base Binary or Dynamic Pitch Mode (not performing a 10-second count, etc), user can press and hold
Background Count Button. After five seconds, a Start Beep sounds and wireless probe will turn off. Note: It takes an
additional five seconds before system displays “
no Linc
”.
If Radionuclide Button is pressed and held for more than one second, user will hear a “start beep” and all radionuclide
indicators will go out. This signifies an Open Window setting. Exit Open Window by pressing and releasing
Radionuclide button (last radionuclide selected is restored).
When in base Binary or Dynamic Pitch Mode (not performing a 10-second count, etc) and a wireless probe battery is
low, Ratio Indicator will show “
Lo
” and Background Count will show “
batt
” for two seconds. A low battery chime
is sounded and No Probe Connected Icon turns on for two seconds, then turns off. Previous values are
restored. Message is seen every 20 seconds.
Artwork No: AW-000639 Revision: A Artwork Status: Released Lifecycle Name: Artwork
PCO No: N/A ECN No: ECN-000214 Release Date: 10/09/2012 Effective Date: 10/09/2012