Acquire a Target Count Ratio
Hold probe stationary over target area. Press and release Background Count Button [14]. Background Count
must be greater than default value (2%) of Count Range selected in order to activate Ratio Calculation
function.
Hold probe stationary over target area. Press and release Target Count Button [11] until a beep sounds; hold
probe steady over target area for 1 to 6 seconds until Count Bar [6] is full and a double Stop Beep sounds.
Collect a Target Count
Target Count Symbol [10] flashes amber.
Target Count Readout [7] displays dashes.
Count Bar [6] fills from bottom to top as data is collected.
Target Count Value
Target Count Readout data [7] and Ratio Readout [9] are displayed for 5 seconds; Target Count Symbol [10]
flashes amber.
Amber Ratio Symbol [12] displayed below Target Count Symbol [10].
Count Bar [6] off.
After 5 seconds, a double Stop Beep sounds and system returns to Dynamic Pitch Mode displaying last
calculated Background Count [8] and current Target Count [7] data.
Acquiring Ten Second Count in Dynamic Pitch Mode
TEN SECOND COUNT
Acquire a Ten Second Count
Hold probe stationary over target area.
Press and hold Target Count Button [11] until beep sounds; then release button.
Hold probe steady for 10 seconds over target area until Count Bar [6] is full and double Stop Beep sounds.
Collect Ten Second Target Count
Target Count Symbol [10] flashes green.
Target Count Readout [7] displays dashes.
Count Bar [6] fills from bottom to top as data is collected.
Artwork No: AW-000639 Revision: A Artwork Status: Released Lifecycle Name: Artwork
PCO No: N/A ECN No: ECN-000214 Release Date: 10/09/2012 Effective Date: 10/09/2012