TARGET COUNT
Acquire Target Count
Hold probe stationary over target area to achieve highest Target Count Readout value [7], highest Count Bar
level, and highest sound.
Press and release Target Count Button [11]; keep probe steady over target area for 1 to 6 seconds until Count
Bar [6] is full and a double Stop Beep sounds.
Collect Target Count
Target Count Symbol [10] flashes green.
Target Count Readout [7] displays dashes.
Count Bar [6] fills from bottom to top as data is collected.
Target Count Value
Target Count Readout data [7] is displayed for 5 seconds; Target Count Symbol (10) flashes green.
Count Bar [6] is off.
After 5 seconds, double Stop Beep sounds and system returns to Dynamic Pitch Mode, displaying last
calculated Background Count [8] and current Count Rate [7].
Acquiring Target Count Ratio in Dynamic Pitch Mode
TARGET COUNT RATIO
Artwork No: AW-000639 Revision: A Artwork Status: Released Lifecycle Name: Artwork
PCO No: N/A ECN No: ECN-000214 Release Date: 10/09/2012 Effective Date: 10/09/2012