GDS Operation Manual
AW-000639
Page
37 of 38
11.
WARNINGS, CAUTIONS AND NOTES
Read this section before operating product.
DEFINITIONS
:
,
or
WARNING: Specific information identifies events when misuse or improper use of
device may cause harm to user or patient or could result in irreparable damage to device
or property.
CAUTION:
Specific information identifies when misuse of equipment may cause malfunction or
produce incorrect readings.
NOTE
: General information to explain or clarify proper operation of product.
Injury
Report injuries immediately to Devicor at
1-877-926-2666
and leave product undisturbed until authorized Devicor
representative can inspect it.
ELECTRICAL:
Verify power cable is secure before each use.
Inspect cables for damage before each use. Use of damaged cables may present an electrical shock hazard.
Disconnect power cord from console to verify fuses. Only use specified rated fuses.
Disconnect power cord from source and disconnect power cord from console before cleaning product.
To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
Maintain proper clearance for access to AC Power Inlet Connector for disconnection from mains power.
Risk of Fire/Explosion
DO NOT USE IN PRESENCE OF FLAMMABLE ANESTHETICS OR EXPLOSIVE GASES.
WARNINGS:
DO NOT GAMMA OR STEAM (AUTOCLAVE) STERILIZE ANY COMPONENT, PART, OR
ACCESSORY.
No modification of this equipment is allowed. Opening or tampering with product may cause damage; void
warranty.
Only properly trained personnel should operate product.
Do not sterilize or immerse in liquid as permanent damage may result.
Use of unauthorized accessories will void warranty; user assumes all liability.
Do not connect unauthorized equipment or accessories to external serial port connector during clinical use. Any person
who connects additional equipment to the signal input or signal output part configures a medical system, and is therefore
responsible for ensuring that the system complies with the systems requirements of IEC 60601-1.
CAUTIONS:
System components should not touch other electrical equipment during use.
Artwork No: AW-000639 Revision: A Artwork Status: Released Lifecycle Name: Artwork
PCO No: N/A ECN No: ECN-000214 Release Date: 10/09/2012 Effective Date: 10/09/2012