Take the short blue tube from the spare parts kit.
Attach it to the metal grommet at the end of the long, blue crystal tube.
Press on the pink O-ring to remove the pink tube attached to the waste container
from the machine.
Attach the pink tube from the waste filter to the other end of the short blue tube from the spare parts kit
which is already connected to the blue tube coming out of the machine.
Return to the front of the machine.
Increase the vacuum pressure by turning the knob all the way to the right.
Turn on the machine.
Let the machine run for one minute.
When the machine stops, remove the short blue tube that you just installed.
Return the short blue tube to the spare parts kit.
Reconnect the long blue tube to the crystal container
Reconnect the pink tube to the pink connector inside the machine.
Run the machine.
The MegaPeel EX should now be receiving crystals.
WHAT IF CRYSTALS ARE NOT COMING OUT OF THE HAND-PIECE? #2
Disconnect the pink and blue tubes from the front of the machine.
Push down on their O-rings.