Maintenance of Diamond Hand-piece
CLEANING AND STERILIZATION OF HAND-PIECE
NOTE
To clean the Diamond hand-piece, first remove the Diamond hand-piece from the tubing
attachment using the small screwdriver found in the Tip Box.
Carefully insert the end of the screwdriver into the space between the Diamond hand-piece and
the metal connector attached to the pink tubing. Use the screwdriver to separate the 2 pieces.
The Diamond hand -piece should then be placed in the Cavicide1, using either a tray or an
ultrasonic unit, making sure they are completely covered and let them soak in CaviCide for 3
minutes at room temperature (68°F/20°C).
STANDARD STEAM STERILIZATION OF HAND-PIECE
Standard steam sterilization is based on your autoclave’s maintenance parameters.
Every sterilizer is different, please consult with your sterilizer’s operating manual for
recommended sterilization times and temperatures.
Steam sterilization should always be used if the surface of the skin may be penetrated.
NOTE
It is not necessary to clean the Diamond hand-piece after each use
General Guidelines: 30 minutes: 250° Fahrenheit at 15psi or 121° Celsius at 1kg/CM2