13. The SmartLite

®

Max Light does generate some heat when in use. If multiple curing cycles in rapid

succession occur, especially in Boost mode, the case may get warm. The unit is equipped with an
automatic thermal safety shut-off and will shut down if the case temperature reaches 41°C. If this occurs,
to avoid skin or mucosa burns, place the unit in the charging cradle or on the counter and wait at least 6
minutes before trying to use the unit again.

2.3 Adverse reactions

1. Prolonged unfiltered exposure to the light source may cause damage to the eye. (See Warnings)
2. Prolonged contact with soft tissue may cause injury or irritation to the tissue. (See Warnings)
3. Medical conditions such as solar urticaria, erythropoietic protoporphyria or cataract surgery may be

aggravated by exposure to emitted light. (See Contraindications, Precautions)

4. Electrical appliances can interfere with some cardiac pacemakers. (See Precautions).

2.4 Storage conditions

Inadequate storage conditions may shorten the useful life and may lead to malfunction of the product.
• Store and transport at temperatures between 0ºC/32ºF - 40°C/104°F
• Use the product at room temperature
• Do not freeze
• Protect from moisture

3. STEP-BY-STEP INSTRUCTIONS

3.1 Initial set-up

NOTE:

Upon receiving the SmartLite

®

Max L.E.D. Curing Light, check the packaging and parts for any possible

damage that may have occurred in transit. If damage is apparent, please contact your authorized
DENTSPLY distributor from which the light was purchased.

Remove the contents from the shipping package. Be careful not to drop or impact the
curing light. Choose a flat, level, sturdy surface to place the charging cradle on. The
SmartLite

®

Max Light is shipped in a “hibernation” mode, and the display will be

blank. To wake the unit, press the On/Off button. (Fig 1.)

Before using for the first time, the SmartLite

®

Max Light must be fully charged. For the

first charge, this process will take about 3 hours. Use only the power supply furnished
with the SmartLite

®

Max Light system. Attempting to use another power supply source

may cause injury or damage and will void the SmartLite

®

Max Light warranty.

Attach the probe to the hand piece as shown in Figure 2.

Plug the connector of the power supply into the back of the charging cradle, or plug it
into the bottom end of the SmartLite

®

Max Light itself, as shown in Figure 3. Plug the

other end of the power supply into the appropriate electrical outlet. As mentioned above,
wait for at least 3 hours to completely charge the light before the first use.

NOTE:

Verify the line plug of the power supply plug is properly rated for your facility, and that all connections
comply with safe practices and all local requirements. Always use unit in conjunction with a properly
grounded outlet.

4

Fig. 1

Fig. 2

OR

Fig. 3

Содержание SmartLite Max

Страница 1: ...DIRECTIONS FOR USE...

Страница 2: ...TABLE OF CONTENTS FOR MULTI LANGUAGE...

Страница 3: ...ower supply can be attached if needed ensuring consistent uninterrupted use the entire day The SmartLite Max Light also features a fully rotating head with a complete 360 range of movement allowing yo...

Страница 4: ...ut Curing Light 12VDC 2 0A Power Input Power Supply 90 240VAC 50 60 Hz Pulse Rate 10 Hz Pulse Duty Cycle 50 Operating Temperature 15 30 C 59 86 F 0 95 RH 1000 8000 meters elevation 2 GENERAL SAFETY NO...

Страница 5: ...emitted from the curing probe As with any bright light source staring directly at the source may cause temporary effects such as after image See Warnings 3 It is recommended that all curing light use...

Страница 6: ...Light check the packaging and parts for any possible damage that may have occurred in transit If damage is apparent please contact your authorized DENTSPLY distributor from which the light was purchas...

Страница 7: ...utput is pulsed 10 times a second giving a net effect of half the power output as the continuous mode Ramp There is a steady increase of the output from 0 to 1200mW cm2 within 4 seconds and then remai...

Страница 8: ...buttons to select the desired settings Since the probe can be rotated 360 it could be used both as a wand style with thumb operation or gun style with index finger operation Always use the settings re...

Страница 9: ...lcohol alcohol containing iodophor chlorine or glutaraldehyde based disinfectant solutions will void the SmartLite Max Light warranty DO NOT immerse unit in water or disinfectant Wipe do not spray sol...

Страница 10: ...me PROBLEM I press Start but no light comes out CAUSE Head is not attached properly SOLUTION Make sure the head is pushed completely into the handle such that the head connector is not visible PROBLEM...

Страница 11: ...oduct DAMAGES ARE LIMITED STRICTLY TO REPLACEMENT OF THE PRODUCT DENTSPLY EXPRESSLY DISCLAIMS LIABILITY FOR INCIDENTAL AND CONSEQUENTIAL DAMAGES RESULTING FROM ITS USE Claims covered by this warranty...

Страница 12: ...ENTSPLY Sankin K K 14 9 Yushima 3 chrome Bunkyo ku Tokyo 113 0034 Japan Tel 81 03 5114 1001 Hong Kong Division of DENTSPLY International 25 F Pacific Link Tower Southmark Tower A 11 Yip Hing Street Wo...