1.4 Contraindications
Use in patients prone to photobiological reactions: Do not use the SmartLite
®
Max L.E.D. Curing Light on
patients with a history of photobiological reactions (including patients with Urticaria solaris or erythropoietic
protoporphyrie) or those currently undergoing treatment with photosensitising pharmaceuticals.
1.5 Technical specifications
PARAMETER
VALUE
Dimension in mm (length x diameter)
240 x 30
Range of Light
377-490nm
Output at the Tip (mW/cm
2
)
1200 (standard)/1400 (boost)
Weight in Grams
220
Goes to Sleep Mode in
10 minutes
Battery Life
At Standard mode, over 200 cycles of 10 seconds
Time for Complete Charging
< 3 hours
Battery
Rechargeable Lithium Ion
Operating Temperature
Unit turns off automatically if the temperature at the probe tip reaches
41°C
Power Input, Curing Light
12VDC @ 2.0A
Power Input, Power Supply
90-240VAC @ 50-60 Hz
Pulse Rate
10 Hz
Pulse Duty Cycle
50%
Operating Temperature
15-30°C (59-86°F)
0-95% RH
-1000 - +8000 meters elevation
2. GENERAL SAFETY NOTES
Be aware of the following general safety notes and the special safety notes in other chapters of these
directions.
SAFETY SYMBOLS
A number of symbols are used both on the device and throughout this manual. It is important that you
understand the meaning of each one, so that when encountered you will be familiar with the type of
information that is being expressed. Obey all safety messages that follow this symbol to avoid possible injury.
Safety alert symbol
This is the safety alert symbol. It is used to alert you to potential personal injury hazards. Obey all
safety messages that follow this symbol to avoid possible injury.
Protect your eyes from the light generated by this product.
Beware of electrical shock or injury.
Type B Applied Part: Protection from electric shock: Type B