5

3.2 Built-in radiometer

The charging cradle incorporates a built-in power meter, allowing you to see how well
the SmartLite

®

Max Light is performing. (Fig. 4)

The power meter is not intended to measure the performance of the PULSE mode.
Select one of the other modes, and set the timer for 10 seconds. Place the head of
the SmartLite

®

Max Light over the sensor and activate the light. The power meter will

illuminate, showing the relative power output. In STANDARD and RAMP modes, two
green bars is acceptable. In the BOOST mode, the power meter will show a blue bar
above the green ones, indicating the BOOST mode is working at a higher power than
the other modes. Yellow, red or orange lights indicate insufficient power output for
curing (see Troubleshooting section).

Inadequate polymerization due to insufficient curing

• Check compatibility of curing light
• Check curing cycle
• Check curing output before each procedure

3.3 Modes

There are 4 different modes to select from based on the procedure and the
requirement. (Fig. 5)

Standard

– The light output is at a steady 1200mW/cm

2

.

Boost

– This mode provides additional power to the LEDs, helping them

generate a higher power output up to 1400mW/cm

2

. Since the LEDs function at such high power

level, only the 5 second curing cycle is provided to avoid premature LED failure. You can use
multiple boost cycles in succession to obtain longer Boost curing.

Pulse

– The light output is pulsed 10 times a second, giving a net effect of half the power output as

the continuous mode.

Ramp

- There is a steady increase of the output from 0 to 1200mW/cm

2

within 4 seconds and then

remains constant.

Timer selection

Curing time can be selected to run at 5, 10, 15 or 20 seconds for all the modes except the Boost mode which
is programmed to run only at 5 seconds.

3.4 Keypad

There are 3 buttons on the key pad for the timer selection, mode selection
and the on/off. (Fig. 6) The on/off button is used to activate and deactivate
the LEDs. Push once to turn on the light and push it again if you need to
turn it off at any time during the curing cycle. The light will default back to
the full time selected if interrupted.

The probe housing the two LEDs connects to the hand piece on top. The two LEDs are confined within the
reflector surface. The probe also houses a heat sink that helps dissipate the heat generated by the LEDs to
avoid the build up of heat that could damage the LEDs.

3.5 Battery status and cord indicator

The battery indicator on the SmartLite

®

Max Light shows the current status of the battery. When

the RED BATTERY LOW INDICATOR blinks, this means the battery has a low charge and urgent
action should be taken to either place the unit back in the charging cradle, or attach the power
cord directly to the handle. The unit may still operate, but the remaining runtime is very close to 0.
If this symbol is continuously on, the battery doesn’t have enough power to run the light and the
unit will not function.

Fig. 4

10

15

5

2

0

Fig. 5

ON/OFF

Timer

Selection

Mode
Selection

Fig. 6

Содержание SmartLite Max

Страница 1: ...DIRECTIONS FOR USE...

Страница 2: ...TABLE OF CONTENTS FOR MULTI LANGUAGE...

Страница 3: ...ower supply can be attached if needed ensuring consistent uninterrupted use the entire day The SmartLite Max Light also features a fully rotating head with a complete 360 range of movement allowing yo...

Страница 4: ...ut Curing Light 12VDC 2 0A Power Input Power Supply 90 240VAC 50 60 Hz Pulse Rate 10 Hz Pulse Duty Cycle 50 Operating Temperature 15 30 C 59 86 F 0 95 RH 1000 8000 meters elevation 2 GENERAL SAFETY NO...

Страница 5: ...emitted from the curing probe As with any bright light source staring directly at the source may cause temporary effects such as after image See Warnings 3 It is recommended that all curing light use...

Страница 6: ...Light check the packaging and parts for any possible damage that may have occurred in transit If damage is apparent please contact your authorized DENTSPLY distributor from which the light was purchas...

Страница 7: ...utput is pulsed 10 times a second giving a net effect of half the power output as the continuous mode Ramp There is a steady increase of the output from 0 to 1200mW cm2 within 4 seconds and then remai...

Страница 8: ...buttons to select the desired settings Since the probe can be rotated 360 it could be used both as a wand style with thumb operation or gun style with index finger operation Always use the settings re...

Страница 9: ...lcohol alcohol containing iodophor chlorine or glutaraldehyde based disinfectant solutions will void the SmartLite Max Light warranty DO NOT immerse unit in water or disinfectant Wipe do not spray sol...

Страница 10: ...me PROBLEM I press Start but no light comes out CAUSE Head is not attached properly SOLUTION Make sure the head is pushed completely into the handle such that the head connector is not visible PROBLEM...

Страница 11: ...oduct DAMAGES ARE LIMITED STRICTLY TO REPLACEMENT OF THE PRODUCT DENTSPLY EXPRESSLY DISCLAIMS LIABILITY FOR INCIDENTAL AND CONSEQUENTIAL DAMAGES RESULTING FROM ITS USE Claims covered by this warranty...

Страница 12: ...ENTSPLY Sankin K K 14 9 Yushima 3 chrome Bunkyo ku Tokyo 113 0034 Japan Tel 81 03 5114 1001 Hong Kong Division of DENTSPLY International 25 F Pacific Link Tower Southmark Tower A 11 Yip Hing Street Wo...