8 Endodontic Mode
Page 23
8
Endodontic Mode
Endodontic Mode Operations
Endodontic Mode (ENDO) allows the user to regulate/change the motor speed, torque, operational direction,
and auto reversing or stop options of the handpiece system. The combination of the torque and Torque-Stop
Mode options in Endodontic Mode allows the user to designate the motor torque limits and direct the motor
into auto reverse once the specified torque limit is reached.
The Midwest
®
E provides 4 default Memory Settings for ratio, speed, torque, and direction (for details see the
default values’ table for ratio, speed, and torque in Table 3 below).
To individually change these settings simply "overwrite" them as described below. This allows one to compile
their own preferences to fit their option needs.
NOTE:
In Endodontic Mode, the light source is set to "ON" for all presets.
Endodontic Mode
ENDO Mode Icon
SPEED Icon
TORQUE-STOP MODE Icon
DIRECTION OF ROTATION
Icon
GEAR RATIO Icon
TORQUE SETTING Icon
ENDO MODE MEMORY
SETTINGS Icon
(+/-) Parameter Icons
Figure 10
– Endodontic Mode Menu Screen
Key
Ratio /
reduction rate
Speed
Torque
Direction of
Operational Rotation
Torque-Stop Mode
E1
5 : 1
300 rpm
1.5 Ncm
Clockwise
Auto Reverse
E2
5 : 1
300 rpm
1.5 Ncm
Clockwise
Auto Reverse
E3
5 : 1
300 rpm
1.5 Ncm
Clockwise
Auto Reverse
E4
5 : 1
300 rpm
1.5 Ncm
Clockwise
Auto Reverse
Table 3
– ENDO Mode Factory Settings
ENDO Mode allows each user to change the following settings:
Gear Ratio
Rotary Speed
Direction of rotation
Torque-Stop Mode (Auto Stop, Auto Reverse, Auto Reverse Forward)
Torque limit