4
Installation and Preparation for Use
Page 12
Connect the Dental Motor
1.
Connect the motor hose to the control unit
Locate the motor hose pin connector on the back panel of the unit. Snap the hose male connector to the
female end on the unit aligning the pins together as shown in the figure below.
Figure 6
– Connect Motor Hose Tubing to the Control Unit
NOTE:
This motor hose can be disconnected from the control unit by pulling the connector away from the
back of the unit for service and replacement.
2.
Connect the Midwest
®
E motor to the motor hose
tube
Lightly coat the O-rings on the motor with a lubricant spray. Align the pins on the motor with the holes on
the motor hose until the two matches; twist the two together to lock as shown in Figure 7 below.
NOTE:
To Measure the air supply coming out of the end of the motor coupler as shown in Figure 8 below.
The value must be at least 10 l/min.
Figure 7
– Connect the Motor to the Supply Tubing
Figure 8
– Air supply measurement