6 Overview of modules and boards
Dentsply Sirona
6.3 Dentist element
Service Manual Sinius / Sinius CS / Sinius TS
Function of the board
The HAC board is integrated in the workstation via CAN and is used for
controlling the brushless dental motors (BL and BL-ISO-S) and for light
control of the turbines.
● Reading in the hose coding
● Controls the rotating field dependent on the rotor position transmitted
by the motor.
● Controls the torque limitation during therapies
● Supply for instrument light (halogen or LED).
● Controls the sensorless and sensor motors
Test 1
(yellow)
Test 3
(yellow)
Boot
(yellow)
Lights up after switching on and remains lit until completion
of the boot process (approximately 60 seconds)
If the board remains in the bootloader after this:
● No launch command has been received from the
download manager, as the software is not consistent
● There is no connection to the HSA.
Claw active
(yellow)
Displays the active instrument holder via flash code.
5V DC
(yellow)
Operating voltage
Debug
(yellow)
Messages from the J416 motor controller
● 5x Programming
● 6x Error
● 7x Warning
Foot control
(yellow)
Lights up when the foot control is pressed.
Connector
X1
> L423 > SDI X1.18 - X1.20 > X32 HSA
X3
> L423 > SDI X1.18 - X1.20 > X32 HSA
X5
> L431 > Sprayvit supply
X7
> L254 > X7 NAA
X8
> L244 > X8 NAJm
X12
> L258 > Instrument holder 2
X13
> L258 > Instrument holder 3
X14
> L258 > Instrument holder 4
X15
> L258 > Instrument holder 5
Components on the board