7 Replacing components
Service Manual Sinius / Sinius CS / Sinius TS
7.4 Hydraulic/pneumatic modules
båÖäáëÜ
7.4
Hydraulic/pneumatic modules
7.4.1
Releasing the pressure in the water unit
1. Close the angle valves for air and water.
2. Switch the treatment center on.
The pressure in the water unit is released.
3. Activate the 3-way syringe/SPRAYVIT E.
The water pressure is reduced.
Tip: Activate the 3-way syringe/SPRAYVIT E several times if
necessary.
NOTICE
Modules are pressurized
Modules that carry water and air may still be pressurized after switching
off the treatment center.
Pressure in the water lines behind the disinfection system is released
after switching off.
➢ Release the pressure in the modules that carry water and air after
switching off.