MAINTENANCE - 7
7
MAINTENANCE
7.1
Periodic maintenance
CAUTION
Maintenance of the device must be carried out by suitably trained personnel, who have read and
understood all the procedures and information given in this instruction manual, particularly in
chapter 2 “Safety information”.
Always wear sterilized disposable latex gloves.
The following table shows the routine maintenance that the operator or maintenance technician must perform regularly. It
indicates the frequency of the maintenance and describes the type of operation to perform.
To see the total number of cycles performed by the sterilizer since the time of installation, use the following button sequence:
press the
TEST
button, then press the
SET
button and then press and hold button
S134
. Pressing the
CUSTOM
button
provides a display of all aborted cycles. Subtracting the number of aborted cycles from the total number of cycles gives the
number of cycles actually run (successfully completed) by the sterilizer. To exit the process, press the
SET
button.
PERIODIC MAINTENANCE TABLE
Frequency
Type of operation
Operation procedure
Daily
Manual cleaning of the sterilization chamber. This
should be done when the chamber is cold.
Cleaning to be done manually
using a cloth moistened with
demineralized water
Daily
Manual cleaning of the rubber gasket on the door.
This should be done when the chamber is cold.
Cleaning to be done manually
using a cloth moistened with
demineralized water
Weekly
Cleaning or replacement of the demineralized water
filter. (code 105320)
See paragraph 7.3
Every 3 weeks
, or
after 60
cycles
or when the display
indicates NEED CLEANING
Cleaning of the sterilization chamber by means
of the periodic cleaning cycle + Cleaning of the
demineralized water filter.
This should be done when the chamber is cold.
See paragraph 7.2
Every 6 months
, or
after
500 cycles
Replacement of the bacteriological filter
(code 021008)
See paragraph 7.4
7.2
Sterilization chamber automatic cleaning cycle
WARNING
Do not use abrasive products.
Do not run the automatic cleaning cycle with the trays in the sterilization chamber.
Clean the surfaces of the sterilization chamber when the device is cold.
NOTE
The device must be switched off in order to open the door. When the maintenance is finished, close
the door and switch off the device to avoid excessive heating.
1. Remove the rack and the trays from inside the sterilization chamber and
wash them with a common neutral detergent. Thoroughly rinse them in
running water, dry them and keep them in a dry place during the automatic
cleaning cycle.