USING THE STERILIZER - 5
To set the operation of the CUSTOM printer from the sterilizer operator panel, proceed as follows:
• simultaneously press the
S134
and
CUSTOM
buttons to enter the menu;
• scroll the menu using the
button, till PRINTER TYPE is displayed;
• press the
SET
button to enter programming;
• use the
button to set PRINTER CUSTOM;
• confirm settings using the
SET
button;
• keep the
POWER
button pressed to return to the initial screen.
5.12.2 Integrated printer (optional)
The model equipped with an integrated thermal printer requires no
additional installation for printing the sterilization cycle.
Replacing the paper
• Open the cover by pressing on the upper tab;
• Remove the empty roll and insert a new roll with the sensitive side of
the paper upwards;
• Use original chemical paper 57 mm in width by Ø 30 mm;
• Close the door, leaving about 5 cm of paper protruding externally.
The report is printed automatically during the cycle, as with the external
printer. If a printed report is not required, simply open the printer door
before starting the cycle.
NOTE:
1. The rear serial interface port is not installed if there is an integrated printer;
2. If any of the parameters exceed the preset limits, the device automatically
goes into an alarm state; the alarm is also indicated on the printed report.
3. If the cycle is properly completed, there is no need to check whether the
values are within the limits.
5.12.3 Connection to a USB log (optional)
The sterilizer can be connected to a device that saves the sterilization reports on a
USB flash drive instead of printing them on paper.
For connection and use, refer to the user manual for the device.
The port is also used for the diagnostics and technical support system through a
dedicated interface, which is supplied to the authorized technical support centres.
The port is protected by a cap to prevent the insertion of foreign objects that could
damage the device. Do not remove it if the device is not in use.
Do not connect devices not specified by the manufacturer.
CAUTION
Do not connect a LAN connection to the service port; this type of connection could cause the
machine's microprocessor to malfunction and void the warranty.
_____CLASS B____
DOMINA PLUS B
Serial no. 001091
Date 14.02.06
Prog. 1
134 C 5’
Cycle 001343
START
Time 13:45:34
ABORTED
Time 13:45:35
ALARM NO. 7
Facsimile of r
ep
or
t with an A
larm