ALARMS - 6
6
ALARMS
6.1
Overview
When the device is switched on and during each sterilization cycle, the characteristic parameters of the various phases of the
cycle are constantly monitored, together with the proper functioning and perfect condition of all the components.
Any anomaly or fault is immediately indicated on the display through warning messages and alarm codes, together with an
acoustic signal.
The following paragraphs contain tables listing the possible warning messages and alarms.
The tables show:
• in the left-hand column, the message or alarm code that appears on the display;
• in the middle column, the cause of the warning or alarm;
• in the right-hand column, the solution to the problem, which in some cases may be resolved by the operator, while in
others technical assistance is necessary.
6.2
List of warning messages
MESSAGE
ON THE SCREEN
CAUSE
SOLUTION
OPEN DOOR
The door was not opened at the end of the
cycle.
START command with door open.
Open the door. Close the door.
FAIL
Failed cycle
See “Alarm code” table.
DRY FAIL
Drying not completed due to manual
intervention (the material was removed before
the end of the drying phase). Sterilization,
however, is completed.
Press the STOP button.
It is possible to run a new cycle.
ADD H2O
Insufficient water level in the loading tank
(appears before starting the cycle)
Top up the main tank.
FULL H2O
The used water tank is full (appears before
starting the cycle)
Empty the recovery tank.
MANU STOP
The cycle has been stopped manually.
Sterilization is not completed.
Dry the chamber, if wet, and restart the cycle.
BLACK OUT
Blackout during the cycle
Press the POWER button to exit.
Check the electrical socket.
Dry the chamber and restart the cycle.
NEED CLEANING
60 cycles completed without an automatic
cleaning cycle.
Run the automatic cleaning cycle.
NEED SERVICE
A year has passed since the installation date or
more than 1500 cycles have been completed
without any maintenance by the technical
support
The warning message disappears as soon as a
cycle is selected, but will appear again at the
next switching on.
Call for a complete check-up by a qualified
technical service; the message will be reset
after the servicing.
NEED INST
Request for the installation procedure
Run the installation procedure.
NEED TEST
A preventive alarm has been detected
See “Alarm code” table.
TEST FAIL
Negative Vacuum test results
Clean the door gasket and repeat the test.
Call for a technical service.