![dental X dxp DOMINA PLUS B Скачать руководство пользователя страница 29](http://html.mh-extra.com/html/dental-x/dxp-domina-plus-b/dxp-domina-plus-b_operators-manual_2483369029.webp)
29
USING THE STERILIZER - 5
5.5
Arranging the material on the trays before sterilization
Follow the guidelines below to obtain optimum effectiveness from the sterilization process and to preserve the life of the
materials and instruments.
NOTE
Place a chemical sterilization indicator on each tray to avoid sterilising the same load twice or using
unsterilized material. When sterilising wrapped material, place the chemical indicator inside one of the
wraps.
Arrange instruments of different metals, previously separated, on different trays;
When sterilising instruments made of metal other than stainless steel, place a sheet of sterilization paper between the
tray and the instrument to avoid direct contact between the two materials;
position cutting instruments so that they do not come into contact during the sterilization cycle; if necessary, isolate
them with gauze or cotton cloth;
arrange receptacles and containers (tubes, cups, glasses, etc.) on their sides or in an inverted position as water
accumulation must be avoided;
do not place more instruments on the trays than they can contain; any overloading must be avoided;
arrange the instruments with sufficient spacing and ensure that they remain separated throughout the sterilization cycle.
position articulated pieces, such as scissors, in an open position;
do not stack the trays one above another or in direct contact with the walls of the chamber; always use the tray support
provided with the sterilizer;
always use the supplied extraction clamp to insert or remove the trays from the sterilization chamber.