![dental X dxp DOMINA PLUS B Скачать руководство пользователя страница 10](http://html.mh-extra.com/html/dental-x/dxp-domina-plus-b/dxp-domina-plus-b_operators-manual_2483369010.webp)
10
SAFETY INFORMATION - 2
2.4
Residual risks
The sterilization process works by means of pressurized steam at high temperature. When removing a load from the
sterilization chamber, always use suitable tools and personal protective equipment for handling the hot racks and tools.
When opening the sterilizer door, particularly during a cycle failure, a small quantity of steam or hot condensate may be
released; open the door with caution.
WARNING
During normal daily use of the device, residual heat risks persist in the areas marked with special
warning signs, as shown in the figure. Avoid direct contact of body parts with these surfaces.
2.5
Bacteriological risks
If the sterilization cycle is not completed, the load, the trays and their restraint system, as well as the inside of the chamber,
should always be considered as potentially contaminated until a subsequent sterilization cycle has been successfully
completed.
The water in the recovery tank should be considered as contaminated, therefore necessary precaution should be taken
when emptying the tank. Check the integrity of the drain hose before using it.
To avoid cross-contamination, wear a new pair of sterile gloves for each task. Take particular care to replace the sterile
gloves when loading or unloading instruments from the sterilization chamber and during maintenance operations.