35
6. All systems continuously running
(Process light on).
A. Film Sensor Bar not seated
properly.
B. Film caught in entry slot.
C. Dirty Film Sensor Bar. ("FILM
PRESENT" light on with no film
under the Bar.
D. Light striking the Film Entry Slot.
E. Faulty Film Sensor Bar.
F. Defective Control Module.
A. Reseat Bar.
B. Remove film.
C. Clean sensors with damp cotton
swab or cloth and warm water.
D. Remove or lower the light source.
E. Call dealer for service. In the
meantime, use manual backup
procedure: Unplug Sensor Bar.
Turn Main Power Switch OFF and
back ON when ready to process
film. Processor should go into the
process mode for about 7 minutes
and then return to standby mode.
F. Call dealer for service.
7. Replenisher pumps continuously
running.
A. Defective Control Module.
A. Call dealer for service.
8. Unit not replenishing.
A. Replenisher Reservoirs empty.
B. Blockage in replenisher dip tubes
or hoses.
C. Pumps clogged with chemical
build-up.
D. Pumps not primed.
E. Defective pumps or electronics.
A. Fill Reservoirs.
B. Check tubes and hoses for kinks
or blockage and clear if necessary.
C. Flush pumps with warm water.
D. Turn Main Power switch OFF and
back ON to prime or call dealer for
service.
E. Call dealer for service.
9. System not activating when film
is inserted (FILM PRESENT
light not coming on).
A. Using processed film (clear or
black).
B. Wrong placement of intraoral films
in the Film Entry Slot.
C. Film Sensor Bar improperly
seated.
D. Dirty Film Sensor Bar.
E. Film Sensor Bar not plugged in.
F. Defective Film Sensor Bar.
G. Mechanical or electronic failure.
A. Use unprocessed film (green).
B. See Figure 23, page 12.
C. Reseat Film Sensor Bar.
D. Clean Film Sensor Bar with damp
cotton swab or cloth and warm
water.
E. Plug Sensor Bar into Control
Module.
F. Call dealer for service. In
the meantime, use manual backup
procedure described in Item 6.
G. Call dealer for service.
VII. Troubleshooting Guide (Cont.)
9000 ONLY
PROBLEM
CAUSE
CURE