22
Figure 10: Attaching Drain
Hose. Place clamp over
hose
and tighten.
Figure 11: Connecting
Hose
from water inlet solenoid
Figure 12: Connecting cable
to light blue receptacle
marked
SOL.
III. Installation
GENERAL
This section describes the procedure for unpacking the processor, its installation and the initial
test procedure.
WARNING: X-Ray radiation can be harmful to patient, technician and dentist. Inad-
equate lead shielding of the darkroom or film storage area may cause undesirable
fogging or exposure of films. Consult your local codes, Health Department or Dental
Equipment Dealer for proper construction of darkroom or placement of film pro-
cessing equipment in the vicinity of X-Ray radiation sources.
UNPACKING
The processor is shipped complete in a single package which contains the following items: two
Wash Pans, a Drain Hose and Clamp, Water Inlet Hose Solenoid Assembly and a Spare Parts Kit.
An instructional Video on operation and maintenance is also provided.
1.
Remove processor and other materials from the carton.
2.
Remove outer covering and the two Access Covers from the processor.
3.
Remove all tape and protective blockers used to secure internal parts during shipment.
Note:
Take care not to damage the drain hose connector which extends from the bot-
tom rear of the processor. (If the unit is a 9000, take care to cut the strap(s) on the
Replenisher Pumps located behind the Washer/Dryer Module and remove
the card
board from under the Fix Pump.)
ASSEMBLY AND INSTALLATION FOR COUNTERTOP
MOUNTING BASE OR PROCESSOR STAND
1.
Place the processor on the countertop, Mounting Base (optional) or Processor Stand
(optional), aligning the rear drain hole with the 3" (7.6 cm) hole in the mounting surface.
2.
Attach the Drain Hose to the Drain Hose Fitting and secure with the Hose Clamp (Figure
10).
Note:
If the optional Processor Mounting Base is being used, level the processor using
the four leveling feet provided.
3.
Connect the solenoid end of the Water Hose Assembly to the water supply.
4.
Connect the other end of the hose to the short hose on the Washer/Dryer Module at the
rear of the processor (Figure 11).
5.
Connect the solenoid electric cable to the receptacle on the Control Module marked
SOL. There is also a light blue I.D. marker on the connector to match the color strip on
top of the module. (See Figure 12).
6.
A convenient "DENT-X Processor Operation" label is provided. Mount this label to an
adjacent surface for easy viewing and review.