The instrument can be powered by the mains using, for example, the stabilized power supply 
SWD10 input 100÷240Vac output 12Vdc – 1000mA (the positive is in the middle).  
 

 

 

                 

The external diameter of power supply connector is 5.5mm, the internal diameter is 2.1mm.   

Warning: The power supply cannot be used as battery charger. 

If the instrument is connected 

to the external power supply, the  

[~]

 symbol is displayed instead the battery symbol

.

 

  

Malfunctioning upon turning on after battery replacement 

After replacing the batteries, the instrument may not restart correctly; in this case, repeat the 
operation.  
After disconnecting the batteries, wait a few minutes in order to allow circuit condensers to 
discharge completely; then reinsert the batteries. 
 

 
9.1

 

W

ARNING ABOUT BATTERY USE

 

 

Batteries should be removed when the instrument is not used for an extended time. 

 

Flat batteries must be replaced immediately. 

 

Avoid loss of liquid from batteries. 

 

Use waterproof and good-quality batteries, if possible alkaline. Sometimes on the market, it 

is possible to find new batteries with an insufficient charge capacity. 

 
 

10.

  

INSTRUMENT

 

STORAGE 

Instrument storage conditions: 

 

Temperature: -25…+65°C. 

 

Humidity: less than 90% RH without condensation. 

 

During storage avoid locations where: 

 

humidity is high; 

 

the instrument may be exposed to direct sunlight; 

 

the instrument may be exposed to a source of high temperature; 

 

the instrument may be exposed to strong vibrations; 

 

the instrument may be exposed to steam, salt or any corrosive gas. 

 

Some parts of the instrument are made of ABS plastic, polycarbonate: do not use any incompatible 
solvent for cleaning. 
 

Содержание HD32.1

Страница 1: ...Operating manual Thermal Microclimate HD32 1 www deltaohm com English Keep for future reference Companies Brands of GHM...

Страница 2: ...ANALYSIS 26 4 1 1 DR Index Draught Risk 27 4 1 2 WBGT Index 28 4 1 3 WCI Index 29 4 1 4 Radiation Temperature Tr 29 4 1 5 The unit of measurement Unit 29 4 1 6 The maximum minimum and average values...

Страница 3: ...xide concentration 94 7 SERIAL INTERFACE AND USB 96 7 1 THE OPERATING PROGRAM A MICROCLIMATE ANALYSIS 96 7 2 THE OPERATING PROGRAM B DISCOMFORT ANALYSIS 98 7 3 THE OPERATING PROGRAM C PHYSICAL QUANTIT...

Страница 4: ...Analysis operating program The machine can be programmed to perform the logging capture of a measurement session You can specifically set a sampling interval per each session Furthermore the auto star...

Страница 5: ...can simultaneously detect the following physical quantities Temperature Relative humidity Illuminance luminance PAR irradiance Wind speed carbon monoxide concentration CO carbon dioxide concentration...

Страница 6: ...HD32 1 6 V2 1 Thermal Microclimate HD32 1...

Страница 7: ...nu confirms the data entered 7 ESC key Allows exiting from the menu or in case of a submenu exiting from the current level display 8 Navigation keys Allows navigation through the menus 9 PRINT key Sta...

Страница 8: ...urements performed with the probes connected the instrument can directly calculate and display the following well being data WBGT index Tu index WCI index Average radiation temperature tr 2 1 1 WBGT I...

Страница 9: ...TEMPERATURE OF 38 C FOR THE PEOPLE BEING EXAMINED In order to calculate the WBGT index the following probes should be connected The natural ventilation wet bulb temperature probe The globe thermomete...

Страница 10: ...m the ground the formula is corrected as follows WCI 13 12 0 6215 ta 11 37 1 5 v1 5 0 16 0 4275 ta 1 5 v1 5 0 16 where v1 5 is the wind speed measured by the instrument at 1 5 m from the ground The fo...

Страница 11: ...exchange with a man the same thermal radiation power exchanged in the real environment In order to evaluate the average radiation temperature you have to measure The globe thermometer temperature the...

Страница 12: ...air temperature or even lower These conditions mainly in buildings with large window surfaces are not particularly healthy as the warm humid air facilitates the development of pathogenic organisms Fr...

Страница 13: ...symmetry 2 2 1 Unsatisfied with the vertical difference of temperature By detecting the temperatures at various heights you can check the presence of a air temperature vertical gradient This gradient...

Страница 14: ...the PDf index unsatisfied with the floor temperature you should refer to the following regulations ISO 7730 2 2 3 Unsatisfied with the radiant asymmetry The radiant temperature asymmetry tpr is the di...

Страница 15: ...n order to calculate the percentage of the unsatisfied with the radiant asymmetry you should refer to the following regulations ISO 7730 2 3 THE OPERATING PROGRAM C PHYSICAL QUANTITIES The HD32 1 Ther...

Страница 16: ...D32 1 main display is divided in three areas The first area displays the battery s charge status and the current time on the first line and the measured quantities arranged on two columns Tnw natural...

Страница 17: ...ing to ISO 7730 The third area displays the F1 F2 and F3 options Please see the SHIFT FNC key in the following paragraph 3 1 3 The Operating Program C Physical Quantities Display The Thermal Microclim...

Страница 18: ...e available units are C and F by pressing F1 again you can return to C By pressing F3 with next on the last line you go to the combined humidity temperature probe complete with SICRAM module The first...

Страница 19: ...ogram of dry air g Kg Specific Humidity calculated AH Grams of vapor in a cubic meter of dry air g m3 Absolute Humidity calculated Pa Partial vapor pressure hPa calculated H Enthalpy J g calculated Td...

Страница 20: ...urements of a combined speed temperature probe complete with SICRAM module are displayed The second line displays the wind speed The third line displays the flow rate measurement In order to get this...

Страница 21: ...line is selected you can change the flow rate unit of measurement The units of measurement for the flow rate are l s liter s m3 s m3 min m3 h ft3 s ft3 min By pressing F1 when the fourth line is selec...

Страница 22: ...select one of the rows with F2 key and then press repeatedly F1 The available units of measurement depend on the type of probe Type of measurement Unit of Measurement Illuminance Phot lux fcd Irradia...

Страница 23: ...module The visualization is the following one The first line of the display indicates the symbol CO indicating that the measurement of carbon monoxide concentration detected by the CO probe with SICRA...

Страница 24: ...and hour minutes seconds in the first line for about 8 seconds Normally the display shows on the left the icon for the battery s charge status on the right hour minutes seconds The battery symbol beco...

Страница 25: ...key Allows exiting from the menu or in case of a submenu exiting from the current level display MEM key Allows starting and ending a logging session the data sending interval must be set in the menu...

Страница 26: ...If you connect multiple probes of the same type only the first recognized probe is accepted the first two if light probes The probes scanning starts fro input 1 to input 8 The barometric pressure sen...

Страница 27: ...risk wbgt WBGT index wet bulb globe temperature wci WCI index wind chill index trad average radiation temperature Tr unit temperature measurement unit data maximum medium average values 4 1 1 DR Index...

Страница 28: ...t and the turbulence intensity value will be displayed 4 1 2 WBGT Index To display the WBGT index proceed as follows after opening the drop down menu with SHIFT FNC 4 Use the arrow keys to select WBGT...

Страница 29: ...lsius F Fahrenheit or K Kelvin degrees as follows 1 Use SHIFT FNC to open the drop down menu 2 Use the arrow keys to select unit 3 Press ENTER to confirm the selected quantity is displayed in the cent...

Страница 30: ...for further details 4 1 8 Start of a new logging session Press MEM to start a Logging session This key starts and stops the logging of a data block to be saved in the instrument s internal memory The...

Страница 31: ...1 m Tf Temperature at floor level P Net radiation DT Radiant asymmetry temperature By pressing SHIFT FNC a drop down menu is shown with the following information unit allows selection of the unit of...

Страница 32: ...isplay using F1 or F2 The F3 key allows choosing to clear Clr the maximum minimum and average data of the captured measurements In order to clear the data select Clr with F3 Another drop down menu wil...

Страница 33: ...s display of the maximum minimum and average values section Allows setting the pipeline section for flow rate calculation Cal CO2 allows starting of calibration procedure for the CO2 probe Cal CO allo...

Страница 34: ...wn menu Use the arrow keys to select data Press ENTER to confirm the selected quantity is displayed in the central line of the display The three quantities max maximum min minimum and avg average are...

Страница 35: ...n menu by pressing SETUP press and ENTER to access the Info menu The following information on the instrument will be displayed Instrument code and operating program firmware date and version serial nu...

Страница 36: ...Use the arrow keys to select the interval duration from 15 seconds to one hour 2 Press ENTER to confirm and return to the Logging menu 3 Press ESC to return to the Logging menu without making any chan...

Страница 37: ...nd will turn off immediately afterwards thus increasing the battery life Once you have accessed the LOGGING submenu previous par use the arrow keys to select Self shut_off mode If the set Log Interval...

Страница 38: ...op time The following message Enter start time will be displayed 1 Use the arrow keys to select the data to be changed year month day and hour minutes seconds 2 Once selected the data will start blink...

Страница 39: ...To cancel the automatic start setting use the Cancel auto start function as illustrated in the following paragraph NOTE The automatic logging session is started even when the instrument is off If it i...

Страница 40: ...ogs the instrument allows printing the files of the captured data Print selected log and deleting all memory Erase ALL logs The instrument can store up to 64 LOG sessions numbered progressively from 0...

Страница 41: ...the page of the log to be printed will be displayed 1 Use the arrow keys to select the log to be printed press F1 and F3 to go to another page 2 once you select a file the acquisition start date and t...

Страница 42: ...directly 5 3 SERIAL MENU SERIAL COMMUNICATION The Serial submenu allows setting the data transfer speed via serial port Baud rate and the record printing interval Print Interval The LOG sessions can b...

Страница 43: ...s ESC to cancel the change and exit the menu item 5 Press ESC over and over to exit from the submenus 6 Press SETUP to exit immediately from the main menu WARNING The communication between instrument...

Страница 44: ...enter the Reset submenu in order to carry out a complete reset of the instrument proceed as follows 1 Press SETUP 2 Use the arrow keys to select Reset 3 Press ENTER You will get the following message...

Страница 45: ...get the following display 5 Use the arrow keys to select the operating program that you want to install 6 Press ENTER to confirm and wait for the chosen program self installation 7 At the end the ins...

Страница 46: ...0 seconds 5 8 CALIBRATE The Calibrate menu is reserved to Technical Support It reports the calibrations and the last calibration performed To access the Calibrate submenu proceed as follows 1 Press SE...

Страница 47: ...strument using the KEY LOCK function and start the LOG session in order to prevent any undesired access by unauthorized personnel To unlock the instrument repeat the steps above Enter the main menu an...

Страница 48: ...cancel 7 The following message will appear 8 Use the arrow keys to enter the new password 9 Press ENTER to confirm or ESC to cancel and go back to the main menu 10 Press SETUP to exit the main menu d...

Страница 49: ...lysis B Discomfort Analysis TP3207 Temperature probe Sensor type Thin film Pt100 Measurement uncertainty Class 1 3 DIN Measurement range 40 C 100 C Connection 4 wires plus SICRAM module Connector 8 po...

Страница 50: ...150 mm according to ISO 7243 ISO 7726 Sensor type Pt100 Measurement uncertainty Class 1 3 DIN Measurement range 10 C 100 C Connection 4 wires plus SICRAM module Connector 8 pole female DIN45326 Cable...

Страница 51: ...hermometer probe 50 mm Sensor type Pt100 Measurement uncertainty Class 1 3 DIN Measurement range 10 C 100 C Connection 4 wires plus SICRAM module Connector 8 pole female DIN45326 Cable length 2 meters...

Страница 52: ...ers Upper probe T95 15 minutes Lower probe T95 4 minutes The TP3227K probe can be used for simultaneous measurement of temperature at 1 10 m and 0 10 m In order to perform simultaneous measurements at...

Страница 53: ...perature measurement at floor level and at ankle height 0 10 m The TP3227PC has priority on the TP3227 1 if both are connected Sensor type Thin film Pt100 Measurement uncertainty Class 1 3 DIN Measure...

Страница 54: ...al discomfort measurement due to vertical temperature gradient Sensor type Thin film Pt100 Measurement uncertainty Class 1 3 DIN Measurement range 10 C 100 C Connection 4 wires plus SICRAM module Conn...

Страница 55: ...with the radiant asymmetry percentage Sensor type Pyranometer NTC Measurement uncertainty NTC 0 15 Typical spectral sensitivity 10 V W m2 Measurement range 0 C 60 C Connection 4 wires plus SICRAM modu...

Страница 56: ...t wire probe Sensor type NTC 10Kohm Measurement uncertainty 0 2 m s 0 1 1 m s 0 3 m s 1 5 m s Measurement range 0 1 5 m s 0 C 80 C AP3203 30 C 30 C AP3203 F Connection 7 wires plus SICRAM module Conne...

Страница 57: ...Sensor type Pt100 Measurement uncertainty Class A Measurement range 4 C 80 C Connection 4 wires plus SICRAM module Connector 8 pole female DIN45326 Cable length 2 meters Cotton wick length 16 cm Tank...

Страница 58: ...indicated below Remove the sensor cap the cap is not screwed Insert the cotton wick previously dipped with distilled water into the temperature probe The cotton wick must protrude from the probe for a...

Страница 59: ...n film Pt100 for temperature Capacity sensor for relative humidity Measurement uncertainty Temperature 1 3 DIN Relative humidity 2 5 Measurement range Temperature 40 C 100 C Relative humidity 0 100 RH...

Страница 60: ...ype Pt100 Measurement uncertainty Class A Measurement range Natural wet Dry temperature 4 C 80 C 30 C 100 C Connection 7 wires plus 2 input SICRAM module Connector 8 pole female DIN45326 Cable length...

Страница 61: ...e 0 5000ppm Resolution 1ppm Connector 8 female poles DIN45326 Cable length 2 metres Response Time T63 2 minutes Temperature Effect 0 2 C CO2 Typical value Atmospheric pressure effect Compensated with...

Страница 62: ...e Usually 5 years in normal environment conditions Long term stability 5 of the measurement year Typical value Calibration At one point on 0ppm Work Relative Temperature Humidity 5 50 C 0 95 RH no con...

Страница 63: ...complete with clamp and fastening screw to support the probes AM32 Rod complete with 2 clamps and fastening screws to support the probes LP 32 F R Support bracket for photometric radiometric probes f...

Страница 64: ...te with head that can host up to 6 measurement probes The same tripod can be used to support the measurement instrument during data capture The arms fitted with suitable clamps for the measurement pro...

Страница 65: ...o come into contact with live parts above 48V This could be extremely dangerous for the instrument as well as for the operator who could be electrocuted Avoid taking measurements in presence of high f...

Страница 66: ...bulb temperature measurement HP3217DM The commissioning is carried out as follows Remove the cover it is not screwed on Insert the braid in the temperature probe the braid has been previously wet with...

Страница 67: ...module The probes are detected during turn on and this cannot be performed when the instrument is already on therefore if a probe is connected and the instrument is on it is necessary to turn it off...

Страница 68: ...sion 50 C 300 C TP473P I Penetration 50 C 400 C TP473P O Penetration 50 C 300 C TP474C O Contact 50 C 300 C TP475A O Air 50 C 250 C TP472I 5 Penetration 50 C 400 C TP472I 10 Penetration 50 C 400 C 0 1...

Страница 69: ...g Grams of vapor in a kilogram of dry air g m3 Grams of vapor in a cubic meter of dry air hPa Partial vapor pressure hPa J g Enthalpy Td Dew point C or F Tw Wet bulb temperature C or F A measurement i...

Страница 70: ...100 RH 40 150 C 1 5 0 85 RH 2 5 85 100 RH T 15 35 C 2 1 5 measure T remaining range 0 3 C Common characteristics Relative humidity Sensor Capacitive Resolution 0 1 RH Temperature drift 20 C 0 02 RH C...

Страница 71: ...0 27 32 78 0 16 75 29 0 12 30 11 28 0 24 32 44 0 14 75 09 0 11 35 11 25 0 22 32 05 0 13 74 87 0 12 40 11 21 0 21 31 60 0 13 74 68 0 13 45 11 16 0 21 31 10 0 13 74 52 0 16 50 11 10 0 22 30 54 0 14 74...

Страница 72: ...series On request the probes of the AP471 series can be fitted with a telescopic rod that eases measurements in areas difficult to reach for example vents The typical applications are wind speed and f...

Страница 73: ...vigation keys to modify the highlighted digit The area comprised must be between 0 0001 m2 1 cm2 and 1 9999 m2 Pressing F1 toggles the display between m2 and inch2 Press Enter to confirm the informati...

Страница 74: ...ure the environment temperature in the range 25 C 80 C the AP471 S4 probe in the range 0 C 80 C The AP472 S modules are calibrated in the factory no calibration is required by the user AP471 S1 AP471...

Страница 75: ...nsor and introduce the probe in the air flow being measured maintaining the arrow at the top of the probe parallel to the flow as indicated in the figures The probe should be maintained orthogonal to...

Страница 76: ...1 m s Air temperature compensation 0 80 C Care and maintenance of the probes The speed sensor of the AP471 S probes is heated and in the presence of gas vapors could trigger a fire or explosion Do not...

Страница 77: ...HD32 1 77 V2 1 Dimensions 85 87 200 min 880 max 13 8 AP471 S1 110 13 8 AP471 S2 85 87 102 200 min 880 max 13 8 AP471 S3 Flexible Part...

Страница 78: ...HD32 1 78 V2 1 MIN 350 MAX 1020 120 AP471 S4...

Страница 79: ...S2 0 5 20 25 80 60 Greater diameters are suitable for flow measurements in the presence of turbulence with medium low air speeds i e at the exit of the ducts Lower diameters are suitable in applicati...

Страница 80: ...easurement following the instructions provided in this chapter Care and maintenance of the probes The probe performance mainly at low speeds largely depends on the very slow friction of the vane on it...

Страница 81: ...HD32 1 81 V2 1 Dimensions AP472 S1 AP472 S2 Unscrew the handle 3 holding the probe body still in the point 1...

Страница 82: ...opic rod AP471S1 23 7 5 6 2 9 Technical information on Vane wind speed measurement probes using SICRAM module AP472 S1 AP472 S2 Type of Measurements Air velocity calculated flow air temperature Air ve...

Страница 83: ...o the input It is sufficient to connect it If the instrument is already on turn it off and back on again in order for the probe to be detected The unit of measurement is determined according to the pr...

Страница 84: ...6 T temperature coefficient 0 05 K Drift after 1 year 1 Working temperature 0 50 C Reference Standard CIE No 69 UNI 11142 LUMINANCE measurement probe LP 471 LUM 2 complete with SICRAM module and equip...

Страница 85: ...2000 10000 Resolution mol m2 s 0 01 0 1 1 Spectral range 400nm 700nm Calibration uncertainty 5 f2 response according to the cosine law 6 f3 linearity 1 f4 instrument reading error 1digit f5 fatigue 0...

Страница 86: ...00 0 1999 9 Resolution W m2 0 1 10 3 0 001 0 01 0 1 Spectral range 400nm 1050nm Calibration uncertainty 5 f2 response according to the cosine law 6 f3 linearity 1 f4 instrument reading error 1digit f5...

Страница 87: ...999 9 Resolution W m2 0 1 10 3 0 001 0 01 0 1 Spectral range 315nm 400nm Peak 360nm Calibration uncertainty 5 f2 response according to the cosine law 6 f3 linearity 1 f4 instrument reading error 1digi...

Страница 88: ...Resolution W m2 0 1 10 3 0 001 0 01 0 1 Spectral range 280nm 315nm Peak 305nm Calibration uncertainty 5 f2 response according to the cosine law 6 f3 linearity 2 f4 instrument reading error 1digit f5 f...

Страница 89: ...Resolution W m2 0 1 10 3 0 001 0 01 0 1 Spectral range 220nm 280nm Peak 260nm Calibration uncertainty 5 f2 response according to the cosine law 6 f3 linearity 1 f4 instrument reading error 1digit f5 f...

Страница 90: ...ality in internal environments and detecting any loss of CO The sensor for the measurement of CO consists of an electro chemical cell with two electrodes CO probe calibration The probe is calibrated b...

Страница 91: ...nd visualized Calibration completed message Press F2 Exit for coming to the measurement 8 The process is finished Zero CO calibration with nitrogen cylinder code MINICAN 12A 1 Switch the instrument on...

Страница 92: ...ration completed message Close the cylinder tap and remove the CO sensor cowling 9 Press F2 Exit for coming back to calibrate 10 Insert the protection 11 The process is finished Replacement of CO sens...

Страница 93: ...ensitivity and confirm with ENTER The screen for the replacement of CO sensor appears 9 With Up and Down arrows set the sensor sensibility value Press ENTER to confirm the instrument comes back to the...

Страница 94: ...pm The calibration has to be done on one point each new calibration cancels the previous one The concentration of carbon dioxide in air is influenced by different factors the human activities companie...

Страница 95: ...ation without modifying the working conditions 8 If the timer reaches the zero the instrument gives an acoustic signal out that confirms that the calibration is finished Note the instrument rejects th...

Страница 96: ...PC the RS232 serial port is automatically disabled and vice versa The instruments are provided with a complete set of commands and data queries to be sent via the PC All the commands transferred to th...

Страница 97: ...time Tw Tg Ta Pr RH Va Tr WBGT i WBGT o WCI LN A00 A01 B02 B03 Print instrument memory map If a section is allocated a number is displayed if it is free 2 points are displayed LFn Log n 0 started on 2...

Страница 98: ...9 5 C Print temperature at head height H1 Tb 22 0 C Print temperature at body height H2 Tn 21 6 C Print temperature of the net radiometer H3 Tk 19 5 C Print temperature at ankles height H4 Tf 19 5 C P...

Страница 99: ...libration date C4 Probe 4 type serial number calibration date C5 Probe 5 type serial number calibration date C6 Probe 6 type serial number calibration date C7 Probe 7 type serial number calibration da...

Страница 100: ...wer off function DISABLE WC0 Setting SELF off WC1 Setting SELF on RA Sample print 0sec Reading of PRINT interval set RL Sample log 30sec Reading of LOG interval set WA Setting PRINT interval is a hexa...

Страница 101: ...instrument or on the PC 7 4 2 The Erase Function clearing the memory To clear the memory use the Erase Log function see par 5 2 5 Log File Manager The instrument starts clearing the internal memory at...

Страница 102: ...esent or is faulty OVFL Overflow appears when the probe detects a value that exceeds the expected measurement range UFL Underflow appears when the probe detects a lower value than the expected measure...

Страница 103: ...correct measurement and turns off Data stored on memory will remain The battery symbol becomes when the external power supply is connected To replace the batteries proceed as follows 1 Switch the ins...

Страница 104: ...mpletely then reinsert the batteries 9 1 WARNING ABOUT BATTERY USE Batteries should be removed when the instrument is not used for an extended time Flat batteries must be replaced immediately Avoid lo...

Страница 105: ...1 105 V2 1 11 MEASUREMENT REPORTS PRINTING Please find below a few examples of reports created with DeltaLog10 software for the different environments 12 1 MODERATE ENVIRONMENT Norm ISO 7730 Norm ISO...

Страница 106: ...HD32 1 106 V2 1 Moderate Environment Norm ISO 7730...

Страница 107: ...HD32 1 107 V2 1 Moderate Environment Norm ISO 7730...

Страница 108: ...HD32 1 108 V2 1 Moderate Environment Norm ISO 7730...

Страница 109: ...HD32 1 109 V2 1 Moderate Environment Norm ISO 7730...

Страница 110: ...HD32 1 110 V2 1 Moderate Environment Norm ISO 7730...

Страница 111: ...HD32 1 111 V2 1 Moderate Environment Norm ISO 7730...

Страница 112: ...HD32 1 112 V2 1 12 2 WARM ENVIRONMENT Norm ISO 7243 Norm ISO 7243...

Страница 113: ...HD32 1 113 V2 1 Warm Environment Norm ISO 7243...

Страница 114: ...HD32 1 114 V2 1 Warm Environment Norm ISO 7243 Norm ISO 7243...

Страница 115: ...HD32 1 115 V2 1 Warm Environment Norm ISO 7243...

Страница 116: ...HD32 1 116 V2 1 Warm Environment Norm ISO 7243...

Страница 117: ...HD32 1 117 V2 1 Warm Environment Norm ISO 7243...

Страница 118: ...HD32 1 118 V2 1 12 3 DISCOMFORT ANALYSIS Norm ISO 7730 Norm ISO 7730...

Страница 119: ...HD32 1 119 V2 1 Discomfort Analysis Norm ISO 7730...

Страница 120: ...HD32 1 120 V2 1 Discomfort Analysis Norm ISO 7730...

Страница 121: ...HD32 1 121 V2 1 Discomfort Analysis Norm ISO 7730...

Страница 122: ...HD32 1 122 V2 1 Discomfort Analysis Norm ISO 7730...

Страница 123: ...HD32 1 123 V2 1 Discomfort Analysis Norm ISO 7730...

Страница 124: ...HD32 1 124 V2 1 Discomfort Analysis Norm ISO 7730...

Страница 125: ...HD32 1 125 V2 1 12 4 HOT ENVIRONMENT Norm ISO 7933 Norm ISO 7933...

Страница 126: ...HD32 1 126 V2 1 Hot Environment Norm ISO 7933...

Страница 127: ...HD32 1 127 V2 1 Hot Environment Norm ISO 7933...

Страница 128: ...HD32 1 128 V2 1 Hot Environment Norm ISO 7933...

Страница 129: ...HD32 1 129 V2 1 Hot Environment Norm ISO 7933...

Страница 130: ...HD32 1 130 V2 1 Hot Environment Norm ISO 7933...

Страница 131: ...HD32 1 131 V2 1 Hot Environment Norm ISO 7933...

Страница 132: ...HD32 1 132 V2 1 12 5 COLD ENVIRONMENT Norm ISO 11079 Norm ISO 11079...

Страница 133: ...HD32 1 133 V2 1 Cold Environment Norm ISO 11079...

Страница 134: ...HD32 1 134 V2 1 Cold Environment Norm ISO 11079...

Страница 135: ...HD32 1 135 V2 1 Cold Environment Norm ISO 11079...

Страница 136: ...HD32 1 136 V2 1 Cold Environment Norm ISO 11079...

Страница 137: ...HD32 1 137 V2 1 Cold Environment Norm ISO 11079...

Страница 138: ...HD32 1 138 V2 1 Cold Environment Norm ISO 11079...

Страница 139: ...HD32 1 139 V2 1 Cold Environment Norm ISO 11079...

Страница 140: ...HD32 1 140 V2 1 12 6 PHYSICAL QUANTITIES...

Страница 141: ...ime 1s Instrument temperature measurement with Pt100 probe Pt100 measuring range 200 650 C Resolution 0 01 C in the range 199 99 C 0 1 C in the remaining range Accuracy 0 01 C in the range 199 99 C 0...

Страница 142: ...400 baud Data bit 8 Parity None Stop bit 1 Flow Control Xon Xoff Serial cable length Max 15 m USB interface Type 1 1 2 0 electrically isolated Memory divided into 64 blocks Memory capacity 67600 recor...

Страница 143: ...Heat Flow PHS Predicted Heat Strain Model Severe Hot ISO 7243 DeltaLog10 Cold Environments Prog A IREQ Required Insulation DLE Limit Exposure Time RT Limit Exposure Time WCI Wind Chill Index Severe Co...

Страница 144: ...3217DM Two sensor probe for natural ventilation wet bulb temperature and dry bulb temperature measurement instead of HP3201 TP3204S and TP3207 The following table shows the required probes for microcl...

Страница 145: ...Temperature WBGT Outdoor Wet Bulb and Globe Thermometer Temperature with solar radiation SWp Sweat Rate Ep Predicted Evaporative Heat Flow 1 Tre Water Loss Dlim tre Dlimloss50 Dlimloss95 PHS _________...

Страница 146: ...body surface and the naked human body surface Skin average temperature Wet skin fraction Unit convective thermal conductivity Unit radiation thermal conductivity Water partial pressure at environment...

Страница 147: ...e measurement net radiometer In the following table are reported the required probes for microclimate indexes measurement The following indexes are calculated using the DeltaLog10 Discomfort Analysis...

Страница 148: ...nloadable from Delta OHM website operating manual Probes support carrying case and cables must be ordered separately Accessories VTRAP32 Tripod complete with 6 input head and 5 probe holders code HD32...

Страница 149: ...standalone temperature probes Pt100 sensor one for floor level temperature measurement 70 mm height 30 mm the other for temperature measurement at ankle height 3 mm height 100 mm Cable length 2 meters...

Страница 150: ...ith 50 cc of distilled water 14 2 PROBES FOR THE OPERATING PROGRAM C PHYSICAL QUANTITIES 14 2 1 Temperature probes complete with SICRAM module TP472I Pt100 sensor immersion probe Stem 3 mm length 300...

Страница 151: ...rectional hot wire telescopic probe with base measuring range 0 1 5m s Cable length 2 meters VANE PROBES AP472 S1 Vane probe with thermocouple K 100 mm Speed from 0 6 to 25 m s temperature from 25 to...

Страница 152: ...5000ppm Cable L 2m MINICAN 12A Nitrogen cylinder for CO2 calibration at 0ppm Volume 20 litres With adjusting valve MINICAN 12A1 Nitrogen cylinder for CO2 calibration at 0ppm Volume 20 litres Without...

Страница 153: ...HD32 1 153 V2 1 NOTES...

Страница 154: ...HD32 1 154 V2 1 NOTES...

Страница 155: ......

Страница 156: ...zed third parties is excluded from the guarantee The instrument shall be returned FREE OF SHIPMENT CHARGES to your dealer The jurisdiction of Padua applies in any dispute The electrical and electronic...

Страница 157: ...e result of the constant development of the product This may produce some differences between the information written in this manual and the instrument you have purchased We cannot completely exclude...

Страница 158: ...GHM GROUP Delta OHM Delta Ohm S r l a socio unico Via Marconi 5 35030 Caselle di Selvazzano Padova ITALY Phone 39 049 8977150 Fax 39 049 635596 www deltaohm com info deltaohm com V2 1 11 01 2019...

Отзывы: