HD2016
- 9 -
V1.4
3.4
SDI-12
CONNECTION
More SDI-12 sensors can be connected in parallel. The distance between a sensor and the ac-
quisition system should not exceed 60 m. Before connecting the instrument to an SDI-12 net-
work containing other sensors, set the address by using the proper SDI-12 command.
Fig. 3.4.1: SDI-12 connection
3.5
C
ONTACT OUTPUT
The voltage-free contact output simulates the pulse output of a tipping bucket rain gauge. The
contact is isolated and is normally open. At 1 minute intervals, the output generates a pulse
train (T
ON
≈
60 ms, T
OFF
≈
60 ms) as a function of the rainfall measured in the previous min-
ute. Each pulse corresponds to an amount of rainfall configurable from 0.001 to 1 mm/pulse
(default=0.2) by using the command
CP0
of the ASCII proprietary protocol.
Fig. 3.5.1 – contact output connection
Power supply
CP18… cable
Other SDI-12
sensor
Data logger
input
Other SDI-12
sensor
Power supply
CP18… cable
contact 1
st
pole
Data logger
input
contact 2
nd
pole