HD2016
- 14 -
V1.4
Heater
The HD2016R rain gauge has a temperature sensor that detects the internal temperature of the instru-
ment (
not the air temperature
). The heater is activated when the internal temperature falls below the
value set with the command CTL (default 4 °C) and is switched off when the internal temperature rises
above the value set with the command CTH (default 10 °C).
Command
Reply
Description
RTH
address
:
temperature
Reads the heater deactivation temperature
(default = 10.0 °C)
CTH
nn.d
address
: &
Sets the heater deactivation temperature to the value
nn.d
(activation temperature
≤
nn.d
≤
10.0 °C)
RTL
address
:
temperature
Reads the heater activation temperature
(default = 4.0 °C)
CTL
nn.d
address
: &
Sets the heater activation temperature to the value
nn.d
(4.0 °C
≤
nn.d
≤
deactivation temperature)
RTD
address
:
state
Reads the heater enabling state:
state
=
Enabled
⇒
heater is used
state
=
Disabled
⇒
heater is not used
CTD
n
address
: &
Enables/disables the use of the heater:
n
=
0
⇒
disable ;
n
=
1
⇒
enable
CHS
address
:
state
&
Changes the activation state of the heater:
state
=
On
⇒
heater has been switched on
state
=
Off
⇒
heater has been switched off
Resolution of contact output
Command
Reply
Description
RP0
address
:
resolution
Reads the resolution, in mm/pulse, of the contact
output (default = 0.200 mm/pulse)
CP0
n.ddd
address
: &
Sets the resolution of the contact output to the value
n.ddd
mm/pulse (max. 1 mm/pulse)
Counters reset
Command
Reply
Description
CS
address
: &
Resets the partial rainfall value
CLR
address
: &
Resets all the counters
It must be run before the total weight counter reaches
the maximum value 999,999,999 mg
RS485 settings
Command
Reply
Description
ROt
address
:
state
Reads the connection state of the termination resistor
of the RS485 output:
state
=
0
⇒
termination resistor not connected
state
=
1
⇒
termination resistor connected
COt
n
address
: &
Connect/disconnect the termination resistor of the
RS485 output:
n
=
0
⇒
disconnect ;
n
=
1
⇒
connect
RMA
address
:
address
address
Reads the Modbus address (default = 1)
CMA
nnn
address
: &
Sets the Modbus address to
nnn
(1…247)
RMB
address
:
baud rate index
MB baud
Reads the Modbus Baud Rate:
Baud rate index
=
0
⇒
9600
Baud rate index
=
1
⇒
19200 (default)
CMB
n
address
: &
Sets the Modbus Baud Rate of index
n
:
n
=
0
⇒
9600 ;
n
=
1
⇒
19200