HD2016
- 27 -
V1.4
10
TECHNICAL CHARACTERISTICS
Power supply
Measuring circuit: 10…15 Vdc
Heater: 12 Vdc
±
10% (
only
HD2016R
)
Consumption
Measuring circuit:
≈
20 mA (1.5 A while discharging, typically for
less than 1 minute with 300 cc and maximum 5 minutes with full
reservoir)
Heater: 90 W (
only
HD2016R
)
Contact output
Isolated NO (Normally Open) contact (R
ON
≈
1 k
Ω
, T
ON
≈
60 ms)
Digital outputs
RS485 with Modbus-RTU or ASCII proprietary protocol
SDI-12
Resolution
Rainfall: 0.001 mm
Weight: 1 mg
Rate: 0.001 mm/h or 0.001 mm/min
Temperature: 0.1 °C
Accuracy
Rainfall
(*)
: ± 0.2 mm (wind speed < 30 m/s)
Temperature: ± 1 °C
Maximum rainfall rate
1000 mm/h
Operating temperature
+4…+60 °C without heating
-20…+60 °C with heating (
only
HD2016R
)
Heater intervention
temperature
+4 °C (
only
HD2016R
)
Capacity
The automatic discharge of the 3000 cc reservoir allows seamless
rainfall measurement.
Protection degree
IP 64
Collector area
400 cm
2
Minimum section of the wires
of the connecting cable
Measuring circuit: 0.7 mm
2
(AWG 19)
Heater: 2.5 mm
2
(AWG 13,
only
HD2016R
)
(*) Very light rainfalls (less than about 10 g, corresponding to approx. 0.2 mm) are not detected because
the water stagnates in the valves and filters, and does not reach the collecting reservoir.