HD2016
- 15 -
V1.4
Command
Reply
Description
RMP
address
:
mode index
MB mode
Reads the Modbus transmission mode:
mode index
=
0
⇒
8N1
mode index
=
1
⇒
8N2
mode index
=
2
⇒
8E1 (default)
mode index
=
3
⇒
8E2
mode index
=
4
⇒
8O1
mode index
=
5
⇒
8O2
(N=no parity, E=even parity, O=odd parity)
CMP
n
address
: &
Sets the Modbus transmission mode:
n
=
0
⇒
8N1 ;
n
=
1
⇒
8N2 ;
n
=
2
⇒
8E1
n
=
3
⇒
8E2 ;
n
=
4
⇒
8O1 ;
n
=
5
⇒
8O2
(N=no parity, E=even parity, O=odd parity)
RMW
address
:
delay index
MB Tx delay
Reads the receiving mode after the Modbus trans-
mission:
delay
index
=
0
⇒
Violate protocol and go in receiving
mode right after Tx
delay
index
=
1
⇒
Respect protocol and wait for 3.5
characters after Tx (default)
CMW
n
address
: &
Sets the receiving mode after the Modbus transmis-
sion:
n
=
0
⇒
Violate protocol and go in receiving mode
right after Tx
n
=
1
⇒
Respect protocol and wait for 3.5 characters
after Tx
Printing the measures
Command
Reply
Description
SR
n
address
:
measured value
Single print of the measure of index n:
n
=0
⇒
Status register content in hexadecimal
n
=2
⇒
Partial rainfall, in µm, from the last reading
command
n
=4
⇒
Total rainfall in µm
n
=6
⇒
Partial rainfall, in µm, from the last reset
command
n
=8
⇒
Average rainfall rate, in µm/min, calculated
in the last minute
n
=10
⇒
Average rainfall rate, in µm/h, calculated in
the last hour
n
=12
⇒
Rain gauge internal temperature in °C
n
=14
⇒
Rainfall total weight (Pmg) in mg
(*)
n
=16
⇒
Rainfall total weight, in mg, considering the
evaporation (Wmg)
(*)
n
=18
⇒
Weight, in mg, of the current reservoir
content (Tmg)
(*)
n
=20
⇒
Partial rainfall, in mils, from the last reading
command
n
=22
⇒
Total rainfall in mils
n
=24
⇒
Partial rainfall, in mils, from the last reset
command
n
=26
⇒
Average rainfall rate, in mils/min, calculated
in the last minute
n
=28
⇒
Average rainfall rate, in mils/h, calculated
in the last hour
n
=30
⇒
Rain gauge internal temperature in °F
(*)
For details on weight measurements, please see
the paragraph “Notes on weight measurements”.
RR
n
address
:
measured value
Continuous print (once per second) of the measure of
index n (see the command SR for the indexes)