HD2016
- 13 -
V1.4
Instrument general information
Command
Reply
Description
P0
address
: &
Ping
G0
HD2016 Model
G2
SN=
number
Serial
number
G3
Firm. Ver.=
version
Firmware
version
G4
Firm. Date=yyyy/mm/dd
Firmware date
GD
F cal:yyyy/mm/dd hh:mm:ss
Factory calibration date
Reservoir discharge
Command
Reply
Description
CD
address
: &
Runs the reservoir discharge cycle
CVL
address
:
state
&
Changes the state of the reservoir discharge valve:
state
=
Opened
⇒
valve has been opened
state
=
Closed
⇒
valve has been closed
COa
address
: &
Activates the discharge after 20 minutes from the
end of the rainfall if the reservoir filling percentage is
greater than or equal to 80% (2400 cc)
COb
address
: &
Activates the discharge after 20 minutes from the
end of the rainfall if the reservoir filling percentage is
greater than or equal to 10% (300 cc)
RDE
address
:
delay
Reads the delay time before running the discharge at
the end of a rainfall (default = 20 minutes)
CDE
nnn
address
: &
Activates the discharge after
nnn
minutes (20
≤
nnn
≤
480) from the end of the rainfall if the reservoir filling
percentage is greater than the set threshold
RTE
address
:
percentage
Reads the reservoir filling percentage that enables the
discharge (default = 10%
⇒
300 cc)
CTE
nn
address
: &
Activates the discharge (after the delay time set) if
the reservoir filling percentage is greater than
nn
%
(1
≤
nn
≤
99)
Note: the discharge takes place normally at the end of a rainfall after the delay time set; if the reservoir
should overfill during a rainfall, the discharge will take place during the rainfall and the upper retention valve
will close to maintain the current rainfall in the upper part of the rain gauge; at the end of the discharge
phase, the rainfall collected in the upper part of the rain gauge will be discharged in the main reservoir and
weighted.
Upper retention valve
Command
Reply
Description
ROv
address
:
state
Reads the enabling state of the upper retention valve:
state
=
0
⇒
valve is not enabled
state
=
1
⇒
valve is enabled
COv
n
address
: &
Enables/disables the use of the upper retention valve:
n
=
0
⇒
disable ;
n
=
1
⇒
enable
CVH
address
:
state
&
Change the state of the upper retention valve:
state
=
Opened
⇒
valve has been opened
state
=
Closed
⇒
valve has been closed