background image

212385 Rev. A

www. s p e c s e l e c t . c o m

Page 4

Product #

A

B

C

Max.

±11mm (

7

/

16

”)

Min.

All 81T201BT/

81T201HW variations*

292mm 

(11½”)

121mm 

(4¾”)

54mm 

(2.13”)

All 81T231BT/

81T231HW variations*

330mm 

(13”)

121mm 

(4¾”)

54mm 

(2.13”)

* Note: -30 models, B = 114mm - 131 mm (4.5” to 5.18”)

Flushometer MUST be paired with a fixture of equivalent flush 

volume.

BEFORE THE FIRST FLUSH (COMPLETE VALVES)

FLUSH BOTH OLD AND NEW PIPES:

It is important to FLUSH and thoroughly CLEAN both new and old water 

piping to ELIMINATE contaminants (eg. scale, sediment, gravel,

cuttings, solder, etc.) from the line. 

Where the water has a sediment content, a PIPE STRAINER in the supply 

line should alleviate that condition and protect working parts of flushometers 

and faucets.

On a NEW INSTALLATION, always flush the valve 4 or 5 times to clean out 

supply line debris. It is also advisable where more than one flushometer 

is installed on a water line, to flush out the water piping through the last 

flushometer outlet of the pipe line.

INSTALL FILTER SCREEN (included):

Note: 

Place filter in stop bore with conical screen facing the water flow. 

Caution:

 Do not push filter in too far. Let filter seat against face of flush valve 

tail. Periodic cleaning of the filter screen with water is recommended 

depending on local water conditions to remove any dirt or sediment. 

See filter screen M&I for additional installation information.

ELIMINATE AIR FROM VALVES:

The FIRST FLUSH should ELIMINATE all air from the TECK flushometer. DO 

NOT ADJUST flushometer based on the results of the first flush.

TO PREVENT WATER HAMMER:

A water hammer arrestor may be installed at the last flushometer and/or at the back of an individual installation. This assures SMOOTHER 

OPERATION of the valves and longer life for the working parts.
HANDS SHOULD BE CLEAN AND FREE OF GREASE AND OIL during any maintenance or handling of electronic housing components to 

prevent possible damage to internal circuitry.

IF LEFT HAND STOP REQUIRED:

• 

Remove the cover by taking out the two cover screws. See Figure 3.

Caution: DO NOT

 remove the four screws (7b) that hold down the 

cap/solenoid assembly

 (6b). 

• 

You will now see the clear electronic compartment with a grey cover.

• 

Now take hold of the compartment and very carefully, slowly lift the compartment about 1/4” and turn it 

180°.

• 

The sensor eye should now be on the other side.

• 

Gently push the compartment back down, aligning the two screw holes on the clear case with the screw 

holes on the brass solenoid cap/solenoid assembly (6b).

• 

Rotate cover gasket (8a) 180° on brass cap.

• 

Check for pinched wires between cover and clear case.

• 

Ensure grey cover is snapped onto clear case and chrome button is properly seated.

• 

Replace the cover and two cover screws - do not overtighten cover screws.

FLUSH VOLUME ADJUSTMENT:

 

3.4 U.S. gallons / 13.0 litres for water closets

 

1.6 U.S. gallons / 6.0 litres for water closets

 

1.27 U.S. gallons / 4.8 litres for water closets

 

0.5 U.S. gallon / 1.9 litres for urinal valves

 

0.125 U.S. gallon / 0.5 litre for urinal valves

The Regulating Screw (061024A, item #6a) may be adjusted, according to job conditions and fixture installed 

to the proper water volume to flush that particular fixture (except -6, -48, -19 and -05 models which are not field adjustable).

NOTE: 

4.8 and 6 Litre water closets or washdown urinals may require the Angle Stop to be set at only ONE TURN OPEN. For a SHORTER flush, 

turn the Regulating Screw (061024A, item #6a) left (counter-clockwise) and right (clockwise) for a LONGER flush (except -6, -48, -19 and 

-05 models which are not field adjustable).

RECOMMENDED WATER SUPPLY:

W/C - minimum flowing pressure: 25 psi (172 kPa), minimum flow rate: 25 gpm (95 lpm)

UR - minimum flowing pressure: 25 psi (172 kPa), minimum flow rate: 8 gpm (30 lpm)

"D"

25mm

(1")

38mm

(1½”)

Cover Tube

1" Copper Inlet Supply

Adapter

Finished

Wall

Of Fixture Spud 

1" Copper Inlet Adapter Installation

 

1.  Measure distance “D” from centre line

     of fixture spud to finished wall.

2.  Cut copper tube 1” shorter than

     dimension “D”.

3.  Push adapter onto copper tube until it

     stops on the shoulder.

4.  Solder adapter to copper tube.

5.  Cut cover tube 1½” shorter than

     dimension “D”.

6.  Slide the wall flange and cover tube

     over the supply pipe.

7.  Thread the union stop onto the adapter,

     and tighten set screw in flange.

Fig. 2

Fig. 3

Left Hand Stop

Right Hand Stop

Содержание TECK II 81T201BT

Страница 1: ... 16 18 20 19 13 14 15 28 28 27 9c 1 Electronic Exposed TECK II Flush Valves with H2Optics Record model number 81T2__ 1__ __A __ __ __ __ Alternate Flush Volumes leave blank if not applicable 6 6 L litre 1 6 gal fixed flush W C only 13 13 L 3 4 gal flush W C Retrofit Kits only 48 4 8 L 1 27 gal fixed flush W C only 19 1 9 L 0 5 gal fixed flush Urinal only 05 0 5 L 0 125 gal fixed flush Urinal only ...

Страница 2: ...5 060082A Adjustable Tail O Ring Package 20 pkg 16 060504A S S Wall Flange Cover Tube 17 061142A 061143A Electronic Water Closet Repair Kit for 3rd Generation E FV Electronic Urinal Repair Kit for 3rd Generation E FV 18 060697A 060694A 060764A 1 Copper Sweat Inlet Adaptor Copper Sweat Inlet Adaptor FIP Inlet Adaptor 19 060843A 060844A 060859A Retrofit Check Stem Unit Capnut Button Complete Seat Wa...

Страница 3: ...42A 060507A UR Adjustable 1 9L 0 5 gal 061346A W C Fixed 4 8L 1 27 gal 061344A 060507A 062025A 061323A 6 0L 1 6 gal 061345A W C Adjustable 13L 3 4 gal 061352A 060507A 6 0L 1 6 gal 061169A 060507A 062025A NOTE Must use all noted components listed in Table 1 to ensure correct flush volumes Fixed models have epoxy in the adjustment screw WaterSense compliant when paired with a WaterSense fixture rate...

Страница 4: ...NOT remove the four screws 7b that hold down the cap solenoid assembly 6b You will now see the clear electronic compartment with a grey cover Now take hold of the compartment and very carefully slowly lift the compartment about 1 4 and turn it 180 The sensor eye should now be on the other side Gently push the compartment back down aligning the two screw holes on the clear case with the screw holes...

Страница 5: ... is off INSTALLATION INSTRUCTIONS BATTERY FLUSHOMETER for Hardwire skip to Step 3 STEP 1 Install battery flushometer to fixture Open inlet stop STEP 2 Follow instructions on yellow static label Figure 4 Note The batteries are already installed and the product is in hibernation mode waiting to be activated After you remove the yellow label you must push the electronic override button three times wi...

Страница 6: ...e electronic override button to advance to the next adjustment feature A blue and red light will be present This mode is currently inactive and is designed for any future options To advance to the 24 Hour Flush press and hold down electronic override button for 5 seconds until the solid blue light is present Release electronic override button you are now in 24 Hour Flush configuration For Urinals ...

Страница 7: ...nap from battery holder 4 Install the refreshed battery holder back into the electronic housing When the new batteries are installed and connected an initial ascending and descending red light sequence will occur 5 Close the battery compartment lid ensuring wires are neatly tucked into compartment and are not pinched between case and cover Test the electronic override button Blue light flashes onc...

Страница 8: ... water closet flush valve requires a minimum water supply of 1 or larger depending on a number of different factors including water pressure PSI available pipe size and length of pipe run number of fixtures per washroom and per building fixture type fixture usage factor elevation of valve above the water main etc We strongly recommend that pipe size calculations be done to insure proper water supp...

Страница 9: ... Making Adjustments to the Electronic Features on page 5 to reduce the sensor range until inadvertent flushing is resolved If this does not resolve the inadvertent flushing proceed to step 4 4 Remove the two cover screws 1 metal cover 2 cap 3 and electronic override button 4 5 Raise the board 7 up out of the box to get access to the adjustment screw 6 6 Using a 1 16 hex key 5 turn the adjustment s...

Страница 10: ...NCIDENTAL AND CONSEQUENTIAL SPECIAL INDIRECT OR PUNITIVE DAMAGES CONNECTED THEREWITH ARE EXCLUDED AND WILL NOT BE PAID BY DELTA FAUCET Some states provinces do not allow limitations on how long an implied warranty lasts or the exclusion or limitation of incidental or consequential damages so the above limitations or exclusions may not apply to you This warranty gives you specific legal rights and ...

Отзывы: