Other data loggers
The connections and power requirements will usually be the
same as for the DL2e.
If you simply want to log the sensor voltages directly, they can
be treated as differential voltage sources of range 0 – 1.1 V DC,
and the data logger should be configured accordingly.
Warning
: we recommend connecting Profile Probes as
differential voltage sensors because they are
powered
sensors.
Although you can measure Profile Probes single-ended, this will
introduce an additional measurement error that depends mainly
on the length of your cable. It may also have undesirable side
effects on the apparent reading from other sensors attached to
the data logger.
You can either convert the data to soil moisture units after
logging, or program your data logger to convert the output
automatically before logging the data, using the information
supplied in the
Conversion to Soil Moisture
section.