To check the connections, select the
Sensors
tab and click on
to see the PR2 readings.
Start logging
When ready click on
to select the
Logger
window and
Start
to start logging…
Collect the data
…later when you want to collect the data, connect to the DL6
and click on
to select the
Dataset
window from which you
can retrieve and display all stored readings.
See the DL6 User Manual for configuration options.
DL2e connection and configuration
The
Profile Probe
is fitted with a screened 8-way connector.
When used with a DL2e this should be connected using the
PRC/w-05 cable, which provides the following connections.
Cable
Function Notes
Colour
Power V+
Red
5-15V DC, PR2/4 80mA, PR2/6 120 mA.
Power 0V is cable screen.
Power 0V
Black
Signal COM Common signal output
Green
Top sensor, 100 mm depth
Yellow
V
out
1
Grey
V
out
2
Brown
V
out
3
White
V
out
4
Blue
V
out
5
For PR2/4, V
out
5 and 6 are not
connected.
Pink
V
out
6
Notes:
1. These connections are not the same as the PR1.
2. The cable screen serves as the power return and is given black
insulation.
3. Do not connect the Power 0V and Signal common at the
logger. This will create reading errors.
14
z
Operation
Profile Probe User Manual
3.0a