HD50
- 34 -
V1.4
Address Type
Holding Register description
9
SW Mixing ratio higher alarm threshold in g/Kg (x10).
10 SW
A
BSOLUTE HUMIDITY
lower alarm threshold in g/m
3
(x10).
11
SW Absolute humidity higher alarm threshold in g/m
3
(x10).
12 SW
W
ET BULB TEMPERATURE
lower alarm threshold in the set measurement unit
(x10).
13 SW
Wet bulb temperature higher alarm threshold in the set measurement unit
(x10).
16 SW
Lower alarm threshold for
TEMPERATURE
with
NTC10K
sensor of channel
2
in the set measurement unit (x10).
17 SW
Higher alarm threshold for temperature with NTC10K sensor of channel 2
in the set measurement unit (x10).
20 SW
I
LLUMINANCE
(low range, models HD50...I...) lower alarm threshold in
lux.
21
SW Illuminance (low range, models HD50...I...) higher alarm threshold in lux
24 SW
A
TMOSPHERIC PRESSURE
lower alarm threshold in the set measurement unit
(the multiplier depends on the set unit).
25 SW
Atmospheric pressure higher alarm threshold in the set measurement unit
(the multiplier depends on the set unit).
32 SW
CO
2
lower alarm threshold in ppm.
33 SW
CO
2
higher alarm threshold in ppm.
50 SW
Lower alarm threshold for
TEMPERATURE
with
NTC10K
sensor of channel
3
in the set measurement unit (x10).
51 SW
Higher alarm threshold for temperature with NTC10K sensor of channel 3
in the set measurement unit (x10).
64 SW
Lower alarm threshold for
TEMPERATURE
with
Pt100
sensor of HP3517E…
probe
in the set measurement unit (x10).
65 SW
Higher alarm threshold for temperature with Pt100 sensor of HP3517E…
probe
in the set measurement unit (x10).
80 SW
I
LLUMINANCE
(high range, models HD50...I2...) lower alarm threshold in
lux.
81 SW
Illuminance (high range, models HD50...I2...) higher alarm threshold in
lux
Measurement alarm thresholds for configurable inputs
1000 +
200 x (
N
-1)
SW Lower alarm threshold for
TEMPERATURE
with
2-wire Pt100
sensor of
channel
N
in the set measurement unit (x10).
1001 +
200 x (
N
-1)
SW Higher alarm threshold for temperature with 2-wire Pt100 sensor of
channel
N
in the set measurement unit (x10).
1002 +
200 x (
N
-1) SW
Lower alarm threshold for
TEMPERATURE
with
3-wire Pt100
sensor of
channel
N
in the set measurement unit (x10).
1003 +
200 x (
N
-1)
SW Higher alarm threshold for temperature with 3-wire Pt100 sensor of
channel
N
in the set measurement unit (x10).
1004 +
200 x (
N
-1)
SW Lower alarm threshold for
TEMPERATURE
with
4-wire Pt100
sensor of
channel
N
in the set measurement unit (x10).
1005 +
200 x (
N
-1) SW
Higher alarm threshold for temperature with 4-wire Pt100 sensor of
channel
N
in the set measurement unit (x10).
1006 +
200 x (
N
-1)
SW Lower alarm threshold for
TEMPERATURE
with
2-wire Pt1000
sensor of
channel
N
in the set measurement unit (x10).
1007 +
200 x (
N
-1)
SW Higher alarm threshold for temperature with 2-wire Pt1000 sensor of
channel
N
in the set measurement unit (x10).
1008 +
200 x (
N
-1) SW
Lower alarm threshold for
TEMPERATURE
with
3-wire Pt1000
sensor of
channel
N
in the set measurement unit (x10).
Содержание HD50 14b7P TC
Страница 52: ...HD50 52 V1 4 NOTES ...
Страница 53: ...HD50 53 V1 4 NOTE ...
Страница 54: ...HD50 54 V1 4 NOTE ...
Страница 55: ......