HD50
- 33 -
V1.4
Address Type
Input Register description
6039 +
200 x (
N
-1)
SW
Resolution of the quantity associated to channel
N
if the channel is config-
ured as mapped 0…50 mV:
…, -2=100, -1=10, 0=1, 1=0.1, 2=0.01, …
6040 +
200 x (
N
-1)
W
Measurement unit of the quantity associated to channel
N
if the channel is
configured as mapped 4…20 mA input: see the table of indexes
6041 +
200 x (
N
-1) SW
Resolution of the quantity associated to channel
N
if the channel is config-
ured as mapped 4…20 mA input:
…, -2=100, -1=10, 0=1, 1=0.1, 2=0.01, …
6042 +
200 x (
N
-1)
W
Measurement unit of the quantity associated to channel
N
if the channel is
configured as mapped potentiometric input: see the table of indexes
6043 +
200 x (
N
-1) SW
Resolution of the quantity associated to channel
N
if the channel is config-
ured as mapped potentiometric input:
…, -2=100, -1=10, 0=1, 1=0.1, 2=0.01, …
6046 +
200 x (
N
-1)
W
Measurement unit of the quantity associated to channel
N
if the channel is
configured as mapped 0…10 V input: see the table of indexes
6047 +
200 x (
N
-1) SW
Resolution of the quantity associated to channel
N
if the channel is config-
ured as mapped 0…10 V input:
…, -2=100, -1=10, 0=1, 1=0.1, 2=0.01, …
6052 +
200 x (
N
-1)
W
Measurement unit of the quantity associated to channel
N
if the channel is
configured as mapped -50…50 mV: see the table of indexes
6053 +
200 x (
N
-1)
SW
Resolution of the quantity associated to channel
N
if the channel is config-
ured as mapped -50…50 mV:
…, -2=100, -1=10, 0=1, 1=0.1, 2=0.01, …
General information
10000
W
Year of last measurement.
10001
W
Month of last measurement.
10002
W
Day of last measurement.
10003
W
Hour of last measurement.
10004
W
Minutes of last measurement.
10005
W
Seconds of last measurement.
10010
W
Time, in seconds, elapsed since the last transmitted packet.
10011
W
RF signal level.
10013 W
Password level for the current connection:
0=no password, 1=user level, 2= administrator level
Holding Registers
- Read/Write parameters
Address Type
Holding Register description
Measurement alarm thresholds
0 SW
Lower alarm threshold for
TEMPERATURE
with
NTC10K
sensor of channel
1
in the set measurement unit (x10).
1 SW
Higher alarm threshold for temperature with NTC10K sensor of channel 1
in the set measurement unit (x10).
2 SW
RH
lower alarm threshold
in % (x10).
3
SW RH higher alarm threshold
in % (x10).
4 SW
D
EW POINT
lower alarm threshold in the set measurement unit (x10).
5
SW Dew point higher alarm threshold in the set measurement unit (x10).
6 SW
P
ARTIAL VAPOR PRESSURE
lower alarm threshold in hPa (x100).
7
SW Partial vapor pressure higher alarm threshold in hPa (x100).
8 SW
M
IXING RATIO
lower alarm threshold in g/Kg (x10).
Содержание HD50 14b7P TC
Страница 52: ...HD50 52 V1 4 NOTES ...
Страница 53: ...HD50 53 V1 4 NOTE ...
Страница 54: ...HD50 54 V1 4 NOTE ...
Страница 55: ......