- 38 -
6.2 CO
2
sensor calibration
CO
2
sensor calibration check can be performed every three/four years.
For a correct sensor calibration, it is crucial to know and respect the physical
phenomena that represent measurement bases: for this reason, we recommend to
thoroughly follow the information below and to perform new calibrations only if in
possession of adequate technical knowledge.
CO
2
sensor can be calibrated in clean air. Each new calibration deletes the previous
one.
To calibrate the sensor, proceed as follows:
1.
Ensure the instrument is actually in clean air.
2.
Power up the instrument by connecting the serial output to your PC USB port by
means of the special RS45 connection cable.
3.
Wait for at least 15 minutes before proceeding.
4.
Start DeltaLog14 program and follow the guided sensor calibration procedure.
Содержание HD45 17A
Страница 10: ... 10 3 1 HD45 connection diagrams ...
Страница 11: ... 11 3 2 HD46 connection diagrams ...
Страница 62: ... 62 Dimensions in mm HD45 series HD46 series ...
Страница 63: ... 63 Fixing holes ...