HD2108/HD2128

 

-   9   -

 

V2.3

 

 

FUNC/ENTER key

 

 
It enables the display and logging of the maximum (MAX), minimum (MIN) and average (AVG) 
value of the measurements captured by the probe connected to the instrument, updating them with 
the acquisition of new samples. The acquisition frequency is once a second. 

In the models with two inputs, the HD2128.1 and HD2128.2

, on activation of the key the two 

display lines represent the MAX, MIN or AVG values for both inputs. 

In the models with one 

input, the HD2108.1 and HD2108.2

, the display shows the current temperature in the main line 

and the maximum, minimum or average temperature in the secondary line. 
The MAX, MIN and AVG measurements remain in the memory until the instrument is on, even 
after exiting the calculation function. To reset the previous values and restart with a new 
measurement session, press FUNC until the message “FUNC CLR” appears, then use the arrows to 
select YES and confirm using ENTER. 

Attention: the data captured using the Record function cannot be transferred to the PC

 
 

 

HOLD/

 key

 

 
It increases the current parameter when used in the menu; when used in measurement mode, it 
freezes the measurement in progress, and upon application of pressure on the key, the message 

HOLD

 appears in the upper side of the display. To return to the current measurement, press the key 

again. 
Upon turning on the instrument, the 

AutoPowerOff

 function can be disabled by holding the MENU 

key down (please see the ON-OFF key description). 
 
 

 

UNIT/UserCAL key

 

 
During measurement allows selection of the unit of measurement for the input temperature. By 
repeatedly pressing the function key, the different units of measurement are displayed in sequence:  
 

1.

 

°C  

Celsius 

degrees 

2.

 

°F  

Fahrenheit 

degrees 

3.

 

°K  

Kelvin 

degrees 

4.

 

mV  

voltage in mV generated by the probe (hot junction) 

5.

 

mV*C   voltage in mV equivalent to the read temperature indicated at 0°C.  

Represents the voltage signal reported by the table illustrating the thermocouple.  

 
This setting changes the information displayed and the immediate print of data (SERIAL key). 

The 

data recorded using the LOG function (

HD2108.2 and HD2128.2

) and sent to the printer or 

PC through the serial port using the SERIAL function, keep the chosen unit of measurement 
and display it. 

 

 

Содержание HD2108.1

Страница 1: ...Operating manual Thermocouple thermometers HD2108 1 HD2108 2 HD2128 1 HD2128 2 www deltaohm com Companies Brands of GHM English Keep for future reference...

Страница 2: ...19 INSTRUMENT STORAGE 20 SERIAL INTERFACE AND USB 21 STORING AND TRANSFERRING DATA TO A PERSONAL COMPUTER 23 THE LOGGING FUNCTION ONLY FOR HD2108 2 AND HD2128 2 23 CLEARING THE MEMORY ONLY FOR HD2108...

Страница 3: ...n transfer the acquired measurements in real time to a PC or to a portable printer The Max Min and Avg function calculate the maximum minimum or average values Other functions include the relative mea...

Страница 4: ...HD2108 HD2128 4 V2 3 HD2108 1 and HD2108 2 Thermocouple Thermometers with one input...

Страница 5: ...MP LOG key for HD2108 2 during normal operation starts and ends the saving of the data in the internal memory in the menu starts the data transfer from the instrument s memory to the PC 11 Mini USB ty...

Страница 6: ...HD2108 HD2128 6 V2 3 HD2128 1 and HD2128 2 Thermocouple Thermometers with two inputs...

Страница 7: ...ng normal operation starts and ends the saving of the data in the internal memory in the menu starts the data transfer from the instrument s memory to the PC 11 Mini USB type B connector for USB 2 0 o...

Страница 8: ...ndary line Each key specific function is described in detail below ON OFF Auto OFF key The instrument is turned on and off using the ON OFF key The turning on enables all display segments for a few se...

Страница 9: ...the PC HOLD key It increases the current parameter when used in the menu when used in measurement mode it freezes the measurement in progress and upon application of pressure on the key the message H...

Страница 10: ...it displays the difference between the current value and that measured on pressing the key The REL message appears on the display press the key again to return to the current measurement When used in...

Страница 11: ...l port is performed each time the key is pressed Recording LOG is performed with one second intervals even if the interval is set to 0 With an interval from 1 to 3600s continuous data transfer is star...

Страница 12: ...interval menu parameter The data logged between a start and subsequent stop represent a block When the logging function is on the LOG indication is displayed the battery symbol blinks and a beep is is...

Страница 13: ...ess SERIAL to end the continuous printing Before starting the printing with SERIAL set the baud rate To do so select the Baud Rate menu item and select the maximum value equal to 38400 baud by using t...

Страница 14: ...d out by inserting the probe in the liquid in which the measurement has to be performed the hot junction is housed in the end part of the probe In the temperature measurement by penetration the probe...

Страница 15: ...rCal In the HD2128 2 model use the arrows to select the input to which the probe being calibrated is connected choosing the upper connector A or the lower one B confirm with ENTER Use the arrows to se...

Страница 16: ...robes must be held perpendicular against the surface Apply oil or heat conductive paste between the surface and the probe in order to improve contact and reduce reading time Whatever you do do not use...

Страница 17: ...that the probe is measuring a value exceeding the measuring range BURN The probe is disconnected or faulty LOG MEM FULL Memory full the instrument cannot store further data the memory space is exhaust...

Страница 18: ...inting started DAY_ day DUMP_END data transfer complete DUMP_In_PROG data transfer in progress ERR error FUNC CLR max min and average values clearing FUNC CLRD max min and average values clearing HOUR...

Страница 19: ...eplace the batteries in order to turn the instrument back on In order to avoid data loss the logging session is ended if the HD2108 2 and HD2128 2 are logging and battery voltage falls below the minim...

Страница 20: ...ies leaking Always use good quality leakproof alkaline batteries Sometimes on the market it is possible to find new batteries with an insufficient charge capacity INSTRUMENT STORAGE Instrument storage...

Страница 21: ...operating system Cable Instrument port PC port Installation of USB drivers HD2110CSNM RS232 MiniDin RS232 9 pole SubD No C 206 RS232 MiniDin USB Yes CP23 USB Mini USB USB No The instrument standard s...

Страница 22: ...le REL function KP Auto power off function ENABLE KQ Auto power off function DISABLE RA Reading of LOG PRINT interval set RP 600 Battery level Resolut 0 01V RUA U C Channel A unit of measurement RUB U...

Страница 23: ...ant it turns on again to capture the new sample and then turns off Data transfer to the PC only HD2108 2 Press the MENU and then the LOG key to transfer to the PC the data contained in the instrument...

Страница 24: ...set from 1 second to 1 hour please see the Print and log interval menu item on page 11 If the print interval is equal to 0 by pressing SERIAL the single data is sent to the connected device If the pri...

Страница 25: ...measurement instrument must be switched off 2 Using the Delta Ohm HD2110CSNM or C 206 cable connect the measuring instrument to the first free RS232C COM or USB serial port of the PC 3 Turn on the ins...

Страница 26: ...perational safety of the instrument can only be guaranteed under the environmental and electrical operating conditions that are in specified in chapter TECHNICAL CHARACTERISTICS Do not use or store th...

Страница 27: ...apter 100 240Vac 12Vdc 1A Security of memorized data Unlimited independently of battery charge conditions Time Date and time Schedule in real time Precision 1min month max drift Measured values storag...

Страница 28: ...o full scale Instrument accuracy Thermocouple K 0 1 C up to 600 C 0 2 C over 600 C Thermocouple J 0 05 C up to 400 C 0 1 C over 400 C Thermocouple T 0 1 C Thermocouple N 0 1 C up to 600 C 0 2 C over 6...

Страница 29: ...C Temperature range C Tolerance C B 600 1700 0 0025 t 600 800 4 800 1700 0 005 t E 40 375 1 5 40 333 2 5 167 40 2 5 375 800 0 004 t 333 900 0 0075 t 200 167 0 015 t J 40 375 1 5 40 333 2 5 375 750 0...

Страница 30: ...D 9 pole female for RS232C C 206 Connection cable 8 pole MiniDin USB type A With integrated RS232 USB converter CP23 Connection cable Mini USB type B USB type A DeltaLog9 Software for transfer and ma...

Страница 31: ......

Страница 32: ...ies is excluded from the guarantee The instrument shall be returned FREE OF SHIPMENT CHARGES to your dealer The jurisdiction of Padua applies in any dispute The electrical and electronic equipment mar...

Страница 33: ...e result of the constant development of the product This may produce some differences between the information written in this manual and the instrument you have purchased We cannot completely exclude...

Страница 34: ...GHM GROUP Delta OHM Delta Ohm S r l a socio unico Via Marconi 5 35030 Caselle di Selvazzano Padova ITALY Phone 39 049 8977150 Fax 39 049 635596 www deltaohm com info deltaohm com V2 3 28 08 2017...

Отзывы: