HD2108/HD2128
- 28 -
V2.3
USB interface - model
HD2108.2
and
HD2128.2
Type
1.1 - 2.0 electrically isolated
Connections
Input for probes
2-pole female polarized standard mignon
connector
RS232 serial interface
8-pole MiniDin connector
USB interface (only
HD2108.2
and
HD2128.2
)
Mini-USB type B connector
Mains adapter (cod.
SWD10
)
2-pole connector (positive at centre)
Measurement of temperature by Instrument
TC measurement range: K
-200 … +1370°C
TC measurement range: J
-100 … +750°C
TC measurement range: T
-200 … +400°C
TC measurement range: N
-200 … +1300°C
TC measurement range: R
+200 … +1480°C
TC measurement range: S
+200 … +1480°C
TC measurement range: B
+200 … +1800°C
TC measurement range: E
-200 … +750°C
Resolution
0.05°C up to 199.95°C
0.1°C from 200.0°C to full scale
Instrument accuracy
Thermocouple K
±0.1°C up to 600°C
±0.2°C over 600°C
Thermocouple J
±0.05°C up to 400°C
±0.1°C over 400°C
Thermocouple T
±0.1°C
Thermocouple N
±0.1°C up to 600°C
±0.2°C over 600°C
Thermocouple R
±0.25°C
Thermocouple S
±0.3°C
Thermocouple B
±0.35°C
Thermocouple E
±0.1°C up to 300°C
±0.15°C over 300°C
The precision only refers to the instrument. Error due to the thermocouple or to the cold
junction reference sensor is not included.
Temperature drift @ 20°C
0.02%/°C
Drift after 1 year
0.1°C/year
Tolerance of the thermocouple probes:
The tolerance of a type of thermocouple corresponds to the maximum acceptable deviation
from the e.m.f. of any thermocouple of that type, with reference junction at 0°C. The
tolerance is expressed in degrees Celsius, preceded by the sign.
The tolerances refer to the operating temperature expected for the thermocouple, in agreement
with the diameter of the thermoelements.