background image

 

HD2108/HD2128

 

-   10   -

 

V2.3

 

 

 

Calibration of the probe and selection of the type of calibration

 

 
Simultaneous pressure on the UNIT/UserCal and FUNC/Enter keys starts the calibration procedure 
of the temperature probe connected to the instrument. Please see the paragraph dedicated to 
calibration on page 14.  
To select the type of calibration (USER=user or FACT= factory) press the UNIT/UserCal and 
FUNC/Enter keys together, then use the arrows to select the desired item, and confirm using 
ENTER.  
 
 

 

TC SEL/ESC key

 

 
In measurement mode, it selects the type of thermocouple K, J, T, N, R, S, B and E. The selected 
thermocouple is shown in quotation marks in the upper line. 
In the menu, the key clears or cancels the active function (ESC). 
 

 

 

REL/

 key

 

 
In measurement mode, it displays the difference between the current value and that measured on 
pressing the key. The 

REL

 message appears on the display; press the key again to return to the 

current measurement. 
When used in the menu, it decreases the current variable value. 
 
 

 

A-B key - only for HD2128.1

 

 
It displays the difference of the temperatures measured by the probes connected to the inputs A and 
B, and indicates ERR if one of the probes is in error (not connected, faulty or overrange). To end 
the function, press the key again. 
 
 

 

A-B/MENU key - only for HD2128.2

 

 
In measurement mode, it displays in the secondary line the difference of the temperatures measured 
by the probes connected to the inputs A and B, and indicates ERR if one of the probes is in error 
(not connected, faulty or overrange). To end the function, press the key again. 
To access the instrument menu,  press simultaneously the A-B/Menu and FUNC/Enter keys. Please 
see the description illustrated below.   
 

Содержание HD2108.1

Страница 1: ...Operating manual Thermocouple thermometers HD2108 1 HD2108 2 HD2128 1 HD2128 2 www deltaohm com Companies Brands of GHM English Keep for future reference...

Страница 2: ...19 INSTRUMENT STORAGE 20 SERIAL INTERFACE AND USB 21 STORING AND TRANSFERRING DATA TO A PERSONAL COMPUTER 23 THE LOGGING FUNCTION ONLY FOR HD2108 2 AND HD2128 2 23 CLEARING THE MEMORY ONLY FOR HD2108...

Страница 3: ...n transfer the acquired measurements in real time to a PC or to a portable printer The Max Min and Avg function calculate the maximum minimum or average values Other functions include the relative mea...

Страница 4: ...HD2108 HD2128 4 V2 3 HD2108 1 and HD2108 2 Thermocouple Thermometers with one input...

Страница 5: ...MP LOG key for HD2108 2 during normal operation starts and ends the saving of the data in the internal memory in the menu starts the data transfer from the instrument s memory to the PC 11 Mini USB ty...

Страница 6: ...HD2108 HD2128 6 V2 3 HD2128 1 and HD2128 2 Thermocouple Thermometers with two inputs...

Страница 7: ...ng normal operation starts and ends the saving of the data in the internal memory in the menu starts the data transfer from the instrument s memory to the PC 11 Mini USB type B connector for USB 2 0 o...

Страница 8: ...ndary line Each key specific function is described in detail below ON OFF Auto OFF key The instrument is turned on and off using the ON OFF key The turning on enables all display segments for a few se...

Страница 9: ...the PC HOLD key It increases the current parameter when used in the menu when used in measurement mode it freezes the measurement in progress and upon application of pressure on the key the message H...

Страница 10: ...it displays the difference between the current value and that measured on pressing the key The REL message appears on the display press the key again to return to the current measurement When used in...

Страница 11: ...l port is performed each time the key is pressed Recording LOG is performed with one second intervals even if the interval is set to 0 With an interval from 1 to 3600s continuous data transfer is star...

Страница 12: ...interval menu parameter The data logged between a start and subsequent stop represent a block When the logging function is on the LOG indication is displayed the battery symbol blinks and a beep is is...

Страница 13: ...ess SERIAL to end the continuous printing Before starting the printing with SERIAL set the baud rate To do so select the Baud Rate menu item and select the maximum value equal to 38400 baud by using t...

Страница 14: ...d out by inserting the probe in the liquid in which the measurement has to be performed the hot junction is housed in the end part of the probe In the temperature measurement by penetration the probe...

Страница 15: ...rCal In the HD2128 2 model use the arrows to select the input to which the probe being calibrated is connected choosing the upper connector A or the lower one B confirm with ENTER Use the arrows to se...

Страница 16: ...robes must be held perpendicular against the surface Apply oil or heat conductive paste between the surface and the probe in order to improve contact and reduce reading time Whatever you do do not use...

Страница 17: ...that the probe is measuring a value exceeding the measuring range BURN The probe is disconnected or faulty LOG MEM FULL Memory full the instrument cannot store further data the memory space is exhaust...

Страница 18: ...inting started DAY_ day DUMP_END data transfer complete DUMP_In_PROG data transfer in progress ERR error FUNC CLR max min and average values clearing FUNC CLRD max min and average values clearing HOUR...

Страница 19: ...eplace the batteries in order to turn the instrument back on In order to avoid data loss the logging session is ended if the HD2108 2 and HD2128 2 are logging and battery voltage falls below the minim...

Страница 20: ...ies leaking Always use good quality leakproof alkaline batteries Sometimes on the market it is possible to find new batteries with an insufficient charge capacity INSTRUMENT STORAGE Instrument storage...

Страница 21: ...operating system Cable Instrument port PC port Installation of USB drivers HD2110CSNM RS232 MiniDin RS232 9 pole SubD No C 206 RS232 MiniDin USB Yes CP23 USB Mini USB USB No The instrument standard s...

Страница 22: ...le REL function KP Auto power off function ENABLE KQ Auto power off function DISABLE RA Reading of LOG PRINT interval set RP 600 Battery level Resolut 0 01V RUA U C Channel A unit of measurement RUB U...

Страница 23: ...ant it turns on again to capture the new sample and then turns off Data transfer to the PC only HD2108 2 Press the MENU and then the LOG key to transfer to the PC the data contained in the instrument...

Страница 24: ...set from 1 second to 1 hour please see the Print and log interval menu item on page 11 If the print interval is equal to 0 by pressing SERIAL the single data is sent to the connected device If the pri...

Страница 25: ...measurement instrument must be switched off 2 Using the Delta Ohm HD2110CSNM or C 206 cable connect the measuring instrument to the first free RS232C COM or USB serial port of the PC 3 Turn on the ins...

Страница 26: ...perational safety of the instrument can only be guaranteed under the environmental and electrical operating conditions that are in specified in chapter TECHNICAL CHARACTERISTICS Do not use or store th...

Страница 27: ...apter 100 240Vac 12Vdc 1A Security of memorized data Unlimited independently of battery charge conditions Time Date and time Schedule in real time Precision 1min month max drift Measured values storag...

Страница 28: ...o full scale Instrument accuracy Thermocouple K 0 1 C up to 600 C 0 2 C over 600 C Thermocouple J 0 05 C up to 400 C 0 1 C over 400 C Thermocouple T 0 1 C Thermocouple N 0 1 C up to 600 C 0 2 C over 6...

Страница 29: ...C Temperature range C Tolerance C B 600 1700 0 0025 t 600 800 4 800 1700 0 005 t E 40 375 1 5 40 333 2 5 167 40 2 5 375 800 0 004 t 333 900 0 0075 t 200 167 0 015 t J 40 375 1 5 40 333 2 5 375 750 0...

Страница 30: ...D 9 pole female for RS232C C 206 Connection cable 8 pole MiniDin USB type A With integrated RS232 USB converter CP23 Connection cable Mini USB type B USB type A DeltaLog9 Software for transfer and ma...

Страница 31: ......

Страница 32: ...ies is excluded from the guarantee The instrument shall be returned FREE OF SHIPMENT CHARGES to your dealer The jurisdiction of Padua applies in any dispute The electrical and electronic equipment mar...

Страница 33: ...e result of the constant development of the product This may produce some differences between the information written in this manual and the instrument you have purchased We cannot completely exclude...

Страница 34: ...GHM GROUP Delta OHM Delta Ohm S r l a socio unico Via Marconi 5 35030 Caselle di Selvazzano Padova ITALY Phone 39 049 8977150 Fax 39 049 635596 www deltaohm com info deltaohm com V2 3 28 08 2017...

Отзывы: