HD2108/HD2128
- 22 -
V2.3
Command
Response
Description
G0
Model HD2128 -2
Instrument model
G1
M=Thermometer 2TC Type K Model description
G2 SN=12345678
Instrument
serial
number
G3 Firm.Ver.=01-00
Firmware
version
G4 Firm.Date=2004/06/15 Firmware
date
G5
cal 0000/00/00 00:00:00
Calibration date and time
G6
Probe=Tc K
Type of probe set
G7
Probe SN=11119999
Probe serial number
G8
Probe cal.=2004/01/12
Probe calibration date
GB
User ID=0000000000000000
User code (set with T2xxxxxxxxxxxxxxxx)
GC
instrument's
heading
LN
&1999
Number of free pages in the flash memory
LD
PRINTOUT OF LOG
Print data logged in flash
LE
&
Erase data in flash memory
K1
PRINTOUT IMMEDIATE
MODE
Immediate printing of data
K0
Stop printing data
K4
&
Start logging data
K5 &
Stop
logging
data
K7
&
Enable REL function
K6
&
Disable REL function
KP &
Auto-power-off
function=ENABLE
KQ &
Auto-power-off
function=DISABLE
RA
& #
Reading of LOG/PRINT interval set
RP
& 600
Battery level (Resolut. 0.01V)
RUA
U= °C
Channel A unit of measurement
RUB U=
°C
Channel B unit of measurement (only HD2128.1 and HD2128.2
models)
WA# &
Setting LOG/PRINT interval.
# is a hexadecimal number 0…D that represents the position of the
interval in the list 0, 1, 5, 10, …, 3600 seconds.
WC0
&
Setting SELF off
WC1
&
Setting SELF on
Command characters are exclusively upper case characters. Once a correct command is entered, the
instrument responds with “&”; when any wrong combination of characters is entered, the
instrument responds with “?”. The instrument response strings end with the sending of the CR
command (Carriage Return). The instrument does not send the LF command (Line Feed).
Before sending commands to the instrument via the serial port, locking the keyboard to avoid
functioning conflicts is recommended: use the P0 command. When complete, restore the keyboard
with the P1 command.