Delta OHM HD2102.1 Скачать руководство пользователя страница 8

 

FUNC/ENTER key

 

 
During normal measurement this enables the display and logging of the maximum (MAX), 
minimum (MIN) and average (AVG) value of the measurements captured by the probe connected to 
the instrument, updating them with the acquisition of new samples. The acquisition frequency is 
once a second. 
The MAX, MIN and AVG measurements remain in the memory until the instrument is on, even 
after exiting the calculation function. To reset the previous values and restart with a new 
measurement session, press FUNC until the message “FUNC CLR” appears, then use the arrows to 
select YES and confirm using ENTER. 
In the menu, the ENTER key confirms the current parameter and then goes to the next one.  
Pressed together with the UNIT-Q/T key, enables the integral Q(t) calculation function. 
Pressed together with the MENU-Clear Q/T, resets the previous integral Q(t) calculation (please see 
the description of the UNIT key). 

 
Attention: the data captured using the Record function cannot be transferred to the PC

 
 

 

HOLD key/

 

 
It increases the current parameter when used in the menu; when used in measurement mode, it 
freezes the measurement in progress, and upon application of pressure on the key, the message 

HOLD

 appears in the upper side of the display. To return to the current measurement, press the 

key again. 
Upon turning on the instrument, the 

AutoPowerOff

 function can be disabled by holding the HOLD 

key down (please see the ON-OFF key description). 
During calculation of the integral Q(t), by pressing the key the displayed value is frozen 

without 

suspending the calculation in progress

 
 

 

UNIT/Q(t) key

 

 
During measurement and integral Q(t) calculation, allows selection of the unit of measurement. 
Repeatedly pressing this key displays the available units of measurement according to the type of 
connected probe, as reported in the following table. 
 

Type of measurement

 

Unit of 

Measurement

 

Integral Q Unit  

of Measurement

 

Illuminance (Phot) 

lux  -  fcd 

lux

s  -  fcd

Irradiance (RAD - UVA - UVB - UVC) 

W/m

2

  - 

μ

W/cm

2

J/m

2

  -  

μ

J/cm

2

PAR 

μ

mol/(m

2

s) 

μ

mol/m

2

Luminance (LUM 2) 

cd/m

2

---- 

Note:

 

for the LP471 LUM 2 luminance probe, the integral calculation is not provided

.

 

-      - 

8

Содержание HD2102.1

Страница 1: ...This situation can cause possible differences comparing this manual with the instrument you bought We regret in advance for any possible mistake in this manual Data drawings and descriptions included...

Страница 2: ...Photo Radiometer HD2102 1 2...

Страница 3: ...en the key is pressed in the menu decreases the current value 9 SERIAL key starts and ends data transfer to the serial communication port 10 MENU CLR Q t key allows access to and exit from the menu If...

Страница 4: ...Photo Radiometer HD2102 2 4...

Страница 5: ...ey during normal operation starts and ends the saving of the measured or integrated data Q T in the internal memory in the menu starts the data transfer from the instrument s memory to the PC 11 MENU...

Страница 6: ...n the menu When exceeded these thresholds cause the instrument to stop the integral calculation The HD2102 2 instrument is a datalogger It memorizes up to 38 000 samples which can be transferred from...

Страница 7: ...detection of the probe connected to the input and sets the instrument ready for normal measurement MIN cm min s 2 3 m m k If during turning on no probes are connected the message PROB ERR is displaye...

Страница 8: ...ction cannot be transferred to the PC HOLD key It increases the current parameter when used in the menu when used in measurement mode it freezes the measurement in progress and upon application of pre...

Страница 9: ...previously been set for the calculation by pressing the FUNC ENTER and UNIT Q T keys When used in the menu it clears or cancels the active function REL key In measurement mode it displays the differe...

Страница 10: ...e moment of sampling and will turn off immediately afterwards thus increasing the battery life Using the arrows select YES and confirm using ENTER in order to enable the automatic turning off select N...

Страница 11: ...d subsequent stop represent a block When the logging function is on the LOG indication is on the display the battery symbol blinks and a beep is issued each time a logging occurs when using an externa...

Страница 12: ...ing on the display upon pressing the SERIAL key is printed the variable associated with the probe connected to the instrument with its relevant unit of measurement or the integral Q t Before starting...

Страница 13: ...nu When exceeded these thresholds cause the instrument to stop the integral calculation The probe is detected during turn on and this cannot be performed when the instrument is already on therefore if...

Страница 14: ...ressing Start Stop Q t in this status to resume the integration press the key again If one or more limits have been enabled in the menu on reaching of the first the calculation stops Pressing FUNC ENT...

Страница 15: ...following table lists all error indications and information displayed by the instrument and supplied to the user in different operating situations Display indications Explanation ERR This appears if t...

Страница 16: ...LIMT umolm2 integration limit in mol m2 LOG In PROG logging in progress LOG MEM FULL memory full LOG_CLRD memory data cleared LOG_STOP logging complete LOG_STRT logging started MIN USE_UNIT_TO_ZERO S...

Страница 17: ...the batteries in order to turn the instrument back on In order to avoid data loss the logging session is ended if the HD2102 2 is logging and battery voltage falls below the minimum operating level T...

Страница 18: ...Always use good quality leakproof alkaline batteries Sometimes on the market it is possible to find new batteries with an insufficient charge capacity INSTRUMENT STORAGE Instrument storage conditions...

Страница 19: ...USB 2 0 connection does not require the setting of parameters The instruments are provided with a complete set of commands and data queries to be sent via the PC All the commands transferred to the i...

Страница 20: ...5 10 3600 seconds WC0 Setting SELF off WC1 Setting SELF on Command characters are exclusively upper case characters Once a correct command is entered the instrument responds with when any wrong combi...

Страница 21: ...mand press the MENU key until the _LOG_DUMP_or_ERAS item is displayed and then press on LOG key During data transfer the display shows the message DUMP to stop the data transfer press ESC on the instr...

Страница 22: ...g SERIAL again NOTES The print out is formatted across 24 columns Some keys are disabled during serial transmission The following keys work ON OFF HOLD FUNC Max Min Avg and LOG Pressing the HOLD REL a...

Страница 23: ...The parameter remains in the memory until replacement of the batteries 4 Launch the DeltaLog9 application and press CONNECT Wait for the connection to occur and follow the indications on the screen F...

Страница 24: ...mplete the device will be detected on each new connection automatically In order to check if the entire operation was successful in CONTROL PANEL double click on SYSTEM Select Device Manager and conne...

Страница 25: ...ed with instrument off 20 A Mains Output mains adapter 9Vdc 250mA Measuring units lux fcd lux s fcd s W m2 W cm2 J m2 J cm2 mol m2 s mol m2 cd m2 Security of memorized data Unlimited independently of...

Страница 26: ...onnector Mains adapter 2 pole connector positive at centre EMC standard regulations Security EN61000 4 2 EN61010 1 level 3 Electrostatic discharge EN61000 4 2 level 3 Electric fast transients EN61000...

Страница 27: ...f6 T Temp coefficient 0 05 K Drift after 1 year 1 Working temperature 0 50 C Reference standard CIE no 69 LUMINANCE measurement probe LP 471 LUM 2 complete with SICRAM module and equipped with the in...

Страница 28: ...9 99 200 0 1999 9 2000 10000 Resolution mol m2 s 0 01 0 1 1 Spectral range 400nm 700nm Calibration uncertainty 5 f 1 response according to the cosine law 6 f3 linearity 1 f4 instrument reading error 1...

Страница 29: ...9 Resolution W m2 0 1 10 3 0 001 0 01 0 1 Spectral range 400nm 1050nm Calibration uncertainty 5 f 1 response according to the cosine law 6 f3 linearity 1 f4 instrument reading error 1digit f5 fatigue...

Страница 30: ...solution W m2 0 1 10 3 0 001 0 01 0 1 Spectral range 315nm 400nm Peak 360nm Calibration uncertainty 5 f 1 response according to the cosine law 6 f3 linearity 1 f4 instrument reading error 1digit f5 fa...

Страница 31: ...on W m2 0 1 10 3 0 001 0 01 0 1 Spectral range 280nm 315nm Peak 305nm Calibration uncertainty 5 f 1 response according to the cosine law 6 f3 linearity 2 f4 instrument reading error 1digit f5 fatigue...

Страница 32: ...on W m2 0 1 10 3 0 001 0 01 0 1 Spectral range 220nm 280nm Peak 260nm Calibration uncertainty 5 f 1 response according to the cosine law 6 f3 linearity 1 f4 instrument reading error 1digit f5 fatigue...

Страница 33: ...otted Line Probe Spectral sensitivity LP471 ERY The probe LP 9021 ERY measures the effective total irradiance W m2 eff according to the UV action curve CEI EN 60335 2 27 A particular type of photodiod...

Страница 34: ...ngle 2 Measurement range 0 1 cd m2 2000 103 cd m2 LP 471 PAR Quantum radiometric probe for the measurement of the photon flow across the chlorophyll range PAR Photosynthetically Active Radiation 400nm...

Страница 35: ...SERIAL INTERFACE AND USB 19 STORING AND TRANSFERRING DATA TO A PERSONAL COMPUTER 21 THE LOGGING FUNCTION ONLY FOR HD2102 2 21 CLEARING THE MEMORY ONLY FOR HD2102 2 21 THE PRINT FUNCTION 22 CONNECTION...

Страница 36: ...m d fectueux Dieser Garantieschein muss der Spedition beigelegt werden wenn das Ger t an das Kundendienstzentrum gesandt wird WICHTIG Die Garantie ist nur g ltig wenn dieser Abschnitt bis ins Einzelne...

Отзывы: