IRRADIANCE measurement probe LP 471 UVA complete with SICRAM module and
equipped with the instrument
Measurement range (W/m
2
):
0.1
⋅
10
-3
… 999.9
⋅
10
-3
1.000…19.999 20.00…199.99
200.0…1999.9
Resolution (W/m
2
):
0.1
⋅
10
-3
0.001 0.01 0.1
Spectral range:
315nm…400nm (Peak 360nm)
Calibration uncertainty:
<5%
f'
1
(response according to the cosine
law):
<6%
f
3
(linearity):
<1%
f
4
(instrument reading error):
±
1digit
f
5
(fatigue):
<0.5%
Drift after 1 year:
<2%
Working temperature:
0…50
°
C
Typical response curve
0
0,1
0,2
0,3
0,4
0,5
0,6
0,7
0,8
0,9
1
280
300
320
340
360
380
400
420
λ
(nm)
R
e
la
tive sp
ectr
a
l r
e
spo
n
se
- -
30