HD2013
- 24 -
V1.5
Command
Instrument reply
Description
Partial amount of rainfall
aC1!
atttn<CR><LF>
with: ttt = number of seconds necessary for the in-
strument to make the measure available (3
characters)
n = number of detected variables (1 character)
Note: ttt = 000 means that datum is immediately
available.
Request of detection of the
partial amount of rainfall
(from the last reset com-
mand).
aD0!
a+n+r…r<CR><LF>
with:
n = number of emptying of the tipping bucket
r…r = partial amount of rainfall in the unit of meas-
urement set in the instrument (mm or inches)
Reads the partial amount
of rainfall (from the last re-
set command).
Amount of rainfall in the current day
aC2!
atttn<CR><LF>
with: ttt = number of seconds necessary for the in-
strument to make the measure available (3
characters)
n = number of detected variables (1 character)
Note: ttt = 000 means that data are immediately
available.
Request of detection of the
amount of rainfall in the cur-
rent day.
aD0!
a+n+r…r<CR><LF>
with:
n = number of emptying of the tipping bucket
r…r = amount of rainfall in the current day in the
unit of measurement set in the instrument
(mm or inches)
Reads the amount of rain-
fall in the current day.
Amount of rainfall in the previous day
aC3!
atttn<CR><LF>
with: ttt = number of seconds necessary for the in-
strument to make the measure available (3
characters)
n = number of detected variables (1 character)
Note: ttt = 000 means that data are immediately
available.
Request of detection of the
amount of rainfall in the
previous day.
aD0!
a+n+r…r<CR><LF>
with:
n = number of emptying of the tipping bucket
r…r = amount of rainfall in the previous day in the
unit of measurement set in the instrument
(mm or inches)
Reads the amount of rain-
fall in the previous day.