72
T
ECHNICAL
S
PECIFICATIONS OF
P
ROBES
E
QUIPPED WITH
SICRAM
M
ODULE
LP 471 PHOT ILLUMINANCE Probe Equipped with SICRAM Module
(Specifications refer to the instrument combined to the module)
Measuring Range (lux):
0.01…199.99
…1999
…19.99
⋅
10
3
…199.9
⋅
10
3
Resolution (lux):
0.01
1
0.01
⋅
10
3
0.1
⋅
10
3
Spectral Response Range:
follows photopic standard curve V(
λ
)
Calibration Uncertainty:
<4%
f'
1
(Follows Photopic Response V(
λ
)):
<8%
f
2
(Cosine Response):
<3%
f
3
(Linearity):
<1%
f
4
(Display Unit Error):
<0.5%
f
5
(Fatigue):
<0.5%
Operating Temperature:
0…50
°
C
LP 471 LUM 2 LUMINANCE Probe Equipped with SICRAM Module
(Specifications refer to the instrument combined to the module)
Measuring Range (cd/m
2
):
0.1…1999
…19.99
⋅
10
3
…199.9
⋅
10
3
…1.999
⋅
10
6
Resolution (cd/m
2
):
0.1 / 1
0.01
⋅
10
3
0.1
⋅
10
3
0.001
⋅
10
6
Field of View:
2
°
Spectral Response Range:
follows photopic standard curve V(
λ
)
Calibration Uncertainty:
<5%
f'
1
(Follows Photopic Response V(
λ
)):
<8%
f
3
(Linearity):
<1%
f
4
(Display Unit Error):
<0.5%
f
5
(Fatigue):
<0.5%
Operating Temperature:
0…50
°
C
Typical Response Curve
0
0,1
0,2
0,3
0,4
0,5
0,6
0,7
0,8
0,9
1
380
430
480
530
580
630
680
730
780
830
λ
(nm)
Relative spectral response