disabled status of the reserved functions, type in the user password and then confirm it with
the <ENTER> key. To quit the function without making any changes, press the clear key
<ESC/CLR>.
1
1
<ESC> exit/cancel
0
1
2
3
4
5
7
9
6
8
1-2) Change password
To change the user password:
A) enable the reserved functions, if this has not been done, by typing in the present password
in point 1) of the Config submenu (see the previous point:
Reserved function lock
)
B) use the function of the
Change password
menu to type in the new password: type the 8
figures and then confirm with the <ENTER> key. To quit the function without making any
changes, press the clear key <ESC/CLR>.
Note: when the instrument is switched on for the first time, after changing the batteries or af-
ter reset, the user password is automatically set at
12345678
.
1
2
<ESC> exit/cancel
12345678
If user level is enabled, current password is proposed
0
1
2
3
4
5
7
9
6
8
1-3) Probe options
If enabled by a password, this function allows you to configure the type of calibration of each
probe present at the instrument inputs. The standard procedure is as follows: the instrument
detects the presence of a probe at one of its inputs and reads its calibration data.
Serial numbers of the probes connected to the instruments
1
3
0
1
2
3
4
5
7
9
6
8
If the probe contains only factory calibration data (indicated with the code “0 factory”), the
instrument will use these parameters.
If user calibration data are also present in the probe (indicated with the code “1 user”), the
user calibration will be used only if it has been carried out with the same instrument and not
with another one.