17
This standard rule may be modified by configuring the instrument so as to use one type of
calibration rather than another.
0) factory
: the probe, connected to the instrument input, will use the factory calibration values
inserted in the probe memory before sale or after recalibration in the factory, even if a user
calibration carried out with the same instrument is present. this choice is useful, for exam-
ple, if there are doubts as to the correctness of the user calibration.
1) user
: the calibration values used are those referring to a user calibration even if they were
obtained with a different instrument. If these are not present, for example because the
probe is new, the instrument will use the factory values.
To modify the type of calibration of a probe, select it, choose the new type of calibration and
then confirm with the <ENTER> key.
In the example below, the probe connected to input A uses the factory calibration and is con-
figured so as use the user calibration.
1
1
<ESC> exit/cancel
Current type of calibration ( factory)
of the probe connected to input A
To quit the function without making any changes, press the clear key <ESC/CLR> instead of
the <ENTER> key.
2) L
OGGING
The Logging heading comprises the settings of the function for storing the data at instrument in-
put, which may be activated by means of the LOG function (key <2/LOG>).
To access each item in the menu, press the key corresponding to the number shown in front of
each function.
2
<ESC> exit/cancel
key 0: log interval
key 1: self shut-off mode
key 2: automatic start and stop time
key 3: cancel automatic start
key 4: log file manager
0
1
2
3
4
5
7
9
6
8
2-0) Log Interval
This is the time interval between two consecutive logging sessions. To set a new interval, on
the menu select the item
Logging
with the key <2/LOG> and then the function
Log interval
with the key <0>; type in the new interval – from 0001 to 3600 – and then confirm with the
<ENTER> key.
To quit the function without making any changes, press the clear key <ESC/CLR>